Electromagnetic Emissions 

Emissions Test 

Compliance 

RF emissions 
EN 60601-1-2:2007+AC:2010, IEC 60601-1-2:2007, CISPR 
11:2009+A1:2010 and IEC 60601-1-2:2014 

Group 1 

RF emissions 
EN 60601-1-2:2007+AC:2010, IEC 60601-1-2:2007, CISPR 
11:2009+A1:2010 and IEC 60601-1-2:2014 

Class B 

Electromagnetic Immunity 

Immunity Test 

IEC 60601 Test Level 

Compliance Level 

Electrostatic 

discharge 

(ESD) 
IEC 61000-4-2 

±2.0  kV,  ±4.0  kV,  ±6.0kV, 
±8.0 kV contact discharge 
±2.0  kV,  ±4.0  kV,  ±8.0  kV, 
±15.0 kV air discharge 

±2.0  kV,  ±4.0  kV,  ±6.0 
kV,  ±8.0  kV  contact 
(56% RH) 
±2.0  kV,  ±4.0  kV,  ±8.0 
kV,  ±15.0  kV  air  (56% 
RH) 

RF  electromagnetic  field 
immunity test 
IEC 61000-4-3 

10 V/m 

10 V/m 

Power frequency magnetic 
fields 
IEC 61000-4-8 

30 A/m 

30 A/m 

 

Warning

1. The TouchCare® transmitter should be avoided to use in an environment with high 
voltage,  high-intensity  magnetic  field,  where  the  intensity  of  EM  DISTURBANCES  is 
high. 

2. Portable RF Communications equipment should be used no closer than 30 cm (12 
inches) to the TouchCare® transmitter. Otherwise, degradation of the performance 
of the transmitter could result. 

3. Use  of  this  equipment  adjacent  to  or  stacked  with  other  equipment  should  be 
avoided, because it could result in improper operation. If such use is necessary, this 
equipment  and  other  equipments  should  be  observed  to  verify  that  they  are 
operating normally. 