19
use. A Transmitter can be used for up to four
sensor sessions.
Insert a New Sensor
Select an Insertion Site
When inserting the senor, choose an area with
adequate subcutaneous fat. This area should stay
flat during normal daily activities without bending
or creasing. The sensor site should be at least 7.5
cm (3 inches) away from an insulin Pump infusion
site or manual injection site. The insulin might
affect Sensor performance, and you might miss a
low or high blood glucose value.
Avoid inserting the Sensor into areas of skin with
scarring, tattoos, or irritation. Do not place the
Sensor in areas that are constrained by clothing,
such as the belt line or waist. Also do not place
the Sensor in areas that involve rigorous
movement during exercise. Avoid sites 5 cm (2
inches) around navel. Avoid areas with excess
hair, or consider shaving the area.
Shown here are the best body areas (shaded) for
Sensor insertion.