MedRx Video Otoscope
Trouble Shooting Guide
External Light Source
Problem
Fuzzy or out of focus
picture
No picture on display
Indicator light switch on
light source not on
No light from probe tip
Picture is dark
Cleaning probe does not
improve picture
Main Cause
Dirty probe tip (dried
cerumen)
Short in twin lead cable.
Light Source not turned
on
Loose wire connection
Power supply not
plugged in
Short in twin lead cable
Light source not on
Bulb burnt out
Switch on back of
camera in wrong position
Light source too low
Probe dropped and
physically damaged
Solution
Carefully clean probe tip
with an alcohol wipe
Replace twin lead cable
Turn light source on
Re-check wire connection
Plug in power supply
Unplug twin cable from
light source; see if light
comes on, if so replace
twin lead cable
Turn on light source
Replace bulb
Move camera shutter
switch to off position
Turn up intensity of light
source
Call MedRx for service
This guide was designed to assist you in diagnosing minor problems.
Should service be needed, please call MedRx at 888-392-1234 for Technical Support
Monday—Friday, EST 9AM to 5PM