ARC-I-MARCI-2 MedRx AVANT ARC Installation Manual
17
Effective Date: 10-21-2016
Recommended separation distances between
Portable and mobile RF communications equipment and the Avant ARC
The Avant ARC is intended to use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Avant ARC can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Avant ARC as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
meters
150 kHz to 80 MHz
d = 1,17
P
×
80 MHz to 800 MHz
d = 1,17
P
×
800 MHz to 2,5 GHz
d = 2,33
P
×
0,01
0,12
0,12
0,233
0,1
0,37
0,37
0,74
1
1,17
1,17
2,33
10
3,7
3,7
7,40
100
11,7
11,7
23,3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.