ARC-I-MARCI-2 MedRx AVANT ARC Installation Manual
15
Effective Date: 10-21-2016
EMC Precautions
The Avant ARC needs special precautions regarding EMC and needs to be installed and put into
service according to the following EMC information.
List of all cables and maximum lengths of cables, transducers and accessories:
Transducer / Accessories
Maximum Cable length
USB Cable
3,0 meters
Insert Earphones
2,0 meters
Bone B-71
2,0 meters
All Headsets
2,0 meters
All Microphones
2,0 meters
REM Probe
2,0 meters
All Speakers
3,0 meters
Warnings!
•
The use of accessories, transducers and cables other than those specified, with the exception of transducers
and cables sold by the manufacturer of the Avant ARC as replacement parts for internal components, may
result in increased emissions or decreased immunity of the Avant ARC.
•
The Avant ARC should not be used adjacent to or stacked with other equipment and if adjacent or stacked use
is necessary, the Avant ARC should be observed to verify normal operation in the configuration in which it will
be used.
•
The Avant ARC may be interfered with by other equipment, even if that other equipment complies with CISPR
emission requirements.
•
The Avant ARC does not have life supporting function
•
Portable and mobile RF communications equipment can affect the Avant ARC.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Avant ARC is intended for use in electromagnetic environment specific below. The customer or the user of the Avant
ARC should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Avant ARC uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Avant ARC is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low -
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Non applicable
Voltage fluctuations / flicker emissions IEC
61000-3-3
Non applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The Avant ARC is intended for use in electromagnetic environment specific below. The customer or the user of the Avant
ARC should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.