Mark 7 Arterion Injection System
11 - 66
•
the programmed volume is delivered and the volume remaining in the syringe is
sufficient to perform another injection, or
•
the operator release the start switch and the volume remaining in the syringe is
sufficient to perform another injection, or
•
the operator presses and release another start switch connected to the system and the
volume remaining in the syringe is sufficient to perform another injection.
The system disarms when:
•
the operator releases the start switch and the volume remaining in the syringe is
insufficient to perform another injection, or
•
the operator presses and releases another start switch and the volume remaining in the
syringe is in-sufficient to perform another injection, or
•
the operator presses the Display Control Unit screen or Injector Head controls.
During the injection, the Injection Indicator displays below the syringe graphic and the Armed
Light flashes.
2.
Repeat step step
to perform additional injections.
3.
Go to
"11.4-Completing an Injection"
11.3.4 Performing a Phased Injection
NOTE:
The term start switch in this section is used to refer to the hand switch, foot switch
or hand controller.
1.
Press and hold the start switch to start the injection, and continue holding until the injection
is complete. The system injects the contrast per the parameters for each phase.
The system stops injecting and disarms when:
•
the programmed volume is delivered, or
•
the operator releases the start switch, or
•
the operator presses and releases another start switch connected to the system, or
•
the operator presses the Display Control Unit screen or Injector Head controls.
During the injection, the Injection Indicator displays below the syringe graphic and the Armed
Light flashes.
2.
"11.4-Completing an Injection"
.
11.3.5 Performing an Injection with Imaging System Interface (ISI)
NOTE:
To use ISI, it must enabled from the
Options
tab and turned on from the
ISI
tab.
The functionality of the Injector hand switch or foot switch and the Imaging System start switch is
determined by the interconnecting cable and the configuration within the Imaging System. The order in
which the systems are initiated determines whether an injection occurs, an X-ray occurs, or neither.
The typical operating scenarios are described in the following tables. Consult with the Field Engineer
for the Imaging System to confirm its internal configuration.
The tables in the subsections below show the Imaging Systems covered in this section and how each
functions with the
Mark 7 Arterion Injection System.
NOTE:
Call Bayer HealthCare Services to determine compatibility of other Imaging
Systems.
NOTE:
The Bayer catalog numbers, for the Imaging Systems listed below, is located on the
cable near the
Mark 7 Arterion
Power Unit.
Содержание Mark 7 Arterion
Страница 1: ...Operation Manual READ BEFORE USING...
Страница 2: ......
Страница 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Страница 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Страница 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Страница 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Страница 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Страница 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Страница 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Страница 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Страница 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Страница 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Страница 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Страница 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Страница 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Страница 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Страница 161: ......