10
1. Clean the nebulizer components and accessories as
described on page 9. Every other treatment day, soak
the components in a solution of 1 part distilled white
vinegar and 3 parts hot tap water for 1 hour.
DO NOT SOAK THE TUBING. You can substitute
commercially available respiratory equipment
disinfectant for the vinegar – follow manufacturer’s
instructions for mixing.
2. Rinse all parts thoroughly with warm tap water and dry
with a clean, lint-free cloth. Discard vinegar solution
when disinfecting is complete.
3. Tubing and accessories must be stored properly
and must be sterilized periodically after use.
The tubing and accessories are for single-patient
use only. Do not share tubing and accessories with
other users.
4. To clean the compressor unit, wipe with a clean,
damp cloth.
DISINFECTING INSTRUCTIONS