Storage
The machine should be stored in a dust and smoke-free area. The case should be wiped with a clean
cloth using a mild detergent solution.
Symbols
Attention, consult accompanying documents Before operating the instrument, please read the
instructions supplied carefully and keep them safe for future reference
Alternating Current
Inlet filter disc
Mains 'On'
Mains 'Off'
Class II Equipment
Type BF Equipment
This product complies with medical devices directive (93/42/EEC), meeting requirements for safety
and product performance.
Tested and approved by
Nemko,
Norway
Page 4:
Troubleshooting
Problem Action
Air compressor will If the switch is on, yet the motor does not function, not operate when 1. Check
mains lead is securely inserted into wall switched on. socket.
2. For machines with a 3 pin plug (UK) remove and check the fuse in the plug.
If it still does not function there is an internal fault. The machine should be returned to the point of
purchase or an authorised person for examination.
Nebuliser chamber
1. Check inlet filter. Replace if necessary.
bubbles or produces
2. Wash nebuliser chamber. Replace if problem little or no mist. still occurs.
3. Check driveline tubing is not kinked or leaking. Time taken to Replace if necessary. nebulise
increases
4. Service compressor. significantly.