4.
When attaching the modified header cleaner
assemblies or the arterial and venous lines to
the Reprocessing Connector make sure that the
connector is firmly attached. This is indicated by a
distinct
CLICK
from the connecting device as it seats
on the Reprocessing Connector.
5.
If the connector is used with the RENACLEAR
Dialyzer Header Cleaning System you may leave the
Reprocessing Connectors attached to the dialyzer.
After the dialyzer is removed from the RENACLEAR,
it can be attached directly to a RENATRON.
6.
After the dialyzer has been reprocessed
remove the Reprocessing Connector from
the dialyzer and clean off any obvious blood
materials.
7.
Disinfect the Reprocessing Connector in
either 1% RENALIN solution or full strength
ACTRIL for a least
30 minutes
before reusing
the connector.
8.
After disinfection and before use, inspect the connector to
ensure that the connector is not damaged and the O-ring
is intact.
NOTE: Simulated-use and PAA exposure testing
indicates that reprocessing connectors are
unaffected through 3 months of exposure and use.
Discard connectors if they show any signs of wear
or if they become more difficult to attach to the
dialyzer, RENATRON or RENACLEAR devices.
MEDIVATORS
®
, ACTRIL
®
, RENALIN
®
, RENATRON
®
and RENACLEAR
®
are registered trademarks of Medivators Inc.
RENACLEAR Modified
Header Cleaner
Assembly
RENATRON Colder
Connector
0
30
15
www.medivators.com
Manufactured in the USA by:
Medivators Inc.
14605 28th Avenue North
Minneapolis, MN 55447 USA
Toll Free: +1.800.444.4729
Medivators BV
Sourethweg 11
6422PC Heerlen
The Netherlands
Tel: +31.45.5.471.471
Cantel Medical Asia/Pacific Pte Ltd
1A International Business Park
#05-01 Singapore 609933
Tel: +65.6227.9698
Cantel Medical Devices (China) Co. Ltd.
Unit 804-805, Innov Tower Block A,
Hongmei Road, Xuhui 200233 Shanghai
Tel: +86 21 60161380
Fax: +86 21 612109136
© 2015 Medivators Inc. 50090-741 Rev E
CLICK