background image

4. 

When attaching the modified header cleaner 
assemblies or the arterial and venous lines to 
the Reprocessing Connector make sure that the 
connector is firmly attached. This is indicated by a 
distinct 

CLICK

 from the connecting device as it seats 

on the Reprocessing Connector.

5. 

If the connector is used with the RENACLEAR 
Dialyzer Header Cleaning System you may leave the 
Reprocessing Connectors attached to the dialyzer. 
After the dialyzer is removed from the RENACLEAR,  
it can be attached directly to a RENATRON.

6. 

After the dialyzer has been reprocessed 
remove the Reprocessing Connector from 
the dialyzer and clean off any obvious blood 
materials.

7. 

Disinfect the Reprocessing Connector in 
either 1% RENALIN solution or full strength 
ACTRIL for a least 

30 minutes 

before reusing 

the connector.

8. 

After disinfection and before use, inspect the connector to  
ensure that the connector is not damaged and the O-ring  
is intact.

NOTE: Simulated-use and PAA exposure testing 
indicates that reprocessing connectors are 
unaffected through 3 months of exposure and use. 
Discard connectors if they show any signs of wear 
or if they become more difficult to attach to the 
dialyzer, RENATRON or RENACLEAR devices.

MEDIVATORS

®

, ACTRIL

® 

, RENALIN

®

, RENATRON

®

 and RENACLEAR

®

 are registered trademarks of Medivators Inc.

RENACLEAR Modified 
Header Cleaner 
Assembly

RENATRON Colder 
Connector

0

30

15

www.medivators.com

Manufactured in the USA by:
Medivators Inc. 
14605 28th Avenue North
Minneapolis, MN 55447 USA
Toll Free:  +1.800.444.4729

Medivators BV
Sourethweg 11
6422PC Heerlen
The Netherlands
Tel:  +31.45.5.471.471

Cantel Medical Asia/Pacific Pte Ltd
1A International Business Park
#05-01 Singapore 609933
Tel:  +65.6227.9698

Cantel Medical Devices (China) Co. Ltd.
Unit 804-805, Innov Tower Block A,
Hongmei Road, Xuhui 200233 Shanghai
Tel: +86 21 60161380
Fax: +86 21 612109136

© 2015 Medivators Inc. 50090-741 Rev E