Patient Temperature 2 Calibration Error The system is unable to calibrate the Patient Temperature 2 channel within ±
1.0 °C when in Manual Control mode.
Patient Temperature 2 channel is made inactive.
33
Water Temperature High
The primary outlet water temperature is above 44°C (111.2 °F).
34
Water Temperature High
The primary outlet water temperature is above 42.5°C (108.5 °F)
35
Water Temperature Low
The primary outlet water temperature is below 3.5°C (38.3 °F).
36
Water Temperature High
The secondary outlet water temperature is above 44°C (111.2 °F).
37
Water Temperature High
The secondary outlet water temperature is above 43°C (109.4 °F).
38
Water Temperature Low
The secondary outlet water temperature is below 3.0°C (37.4 °F)
40
Unable to Maintain Stable Water
Temperature
In Manual Control mode, the system is unable to control the water tempera-
ture within 1.0 °C/°F of the water target after 25 min. in the current mode or
since last change to water target temperature.
41
Low Internal Flow
Insufficient internal flow during system priming or pre-conditioning.
43
User Settings Not Saved
The user settings are invalid and are unable to be saved. The saved system
default settings are restored.
44
Invalid System Log Entry
One or more of the entries into the system event log is invalid.
The system event log is used by clinical engineering personnel for product
service.
This issue does not affect the performance of the system to deliver patient
therapy.
45
AC Power Lost
The AC power was lost while the power switch was in the On position.
46
Control Panel Communication
The control panel is not communicating with the system.
47
Control Panel Communication
The control panel is not communicating with the system.
48
Patient Temperature Out Invalid
The Patient Temperature Out calibration data in non-volatile memory is
invalid.
50
Patient Temperature 1 Erratic
Patient Temperature 1 discontinuity. There has been a significant change in
patient temperature over the past 8 minutes.
51
Patient Temperature 1 Below Control
Range
Patient Temperature 1 is less than 31°C (87.8°F) while in a patient control
mode (e.g. Control Patient, Cool Patient, or Rewarm Patient).
52
Extended Period of Cold Water
The circulating water temperature has been below 10°C (50°F) for 8 of the
previous 10 hours.
The alert will recur after 1 hour if the condition continues. After the device
has issued 11 extended cold water exposure alerts, it will issue a prolonged
cold water exposure alarm.
Extended periods of cold water delivery may increase the risk for skin injury.
Assess patient’s skin underneath the ArcticGel Pads.
53
Prolonged Cold Water Exposure
The circulating water temperature has been below 10°C (50°F) for a pro-
longed period of time. The extend period of cold water alert has been issued
11 times. The alert was first issued after the system sensed that the water
temperature was below 10°C (50°C) for 8 of 10 hours. The alert was then
issued an additional 10 times every 1 hour because the situation was not
resolved.
Prolonged cold water exposure may increase the risk for skin injury. Assess
patient’s skin underneath the ArcticGel Pads.
60
Non-Recoverable System Error
Control processor and Monitor processor start up synchronization fault.
61
Non-Recoverable System Error
Control processor parameter memory fault.
62
Non-Recoverable System Error
Monitor processor parameter memory fault.
64
Non-Recoverable System Error
Unable to enable pump power (Control processor).