SPECIFICATIONS
microVENT
Classic, Airmix,
Adult / Child
microVENT
Classic,
Adult/Child
microVENT
World, Airmix,
Adult/Child
microVENT
World,
Adult/Child
microVENT
CPR
Part Number-
Advanced
model (UK
specification)
670-0060-00
670-0010-00
670-0531-00
670-0506-00
N/A
Part Number-
Standard
model (UK
specification)
670-0061-00
670-0009-00
670-0539-00
670-0511-00
670-
0483-00
Patient
population
range
Adult
Adult
Adult
Adult
Adult
Child above 20
kg
Child above 20
kg
Child above 10
kg
Child above 10
kg
Child
above10
kg
Automatic
operation
Time cycled.
Gas powered.
(Patient assist
synchronisation
fitted on
Advanced
models)
Time cycled.
Gas powered.
(Patient assist
synchronisation
fitted on
Advanced
models)
Time cycled.
Gas powered.
(Patient assist
synchronisation
fitted on
Advanced
models)
Time cycled.
Gas powered.
(Patient assist
synchronisation
fitted on
Advanced
models)
Time
cycled.
Gas
powered
Automatic
flow rate
(L/min)
43.2 to 21.6
43.2 to 21.6
36 to 11.25
36 to 11.25
36 to 9
Automatic
tidal volume
(L)
1.2 to 0.3
1.2 to 0.3
1.0 to 0.15
1.0 to 0.15
0.6 to
0.15
Automatic
oxygen
concentration
V/V
100% or
50%(nominal)
100%
100% or
50%(nominal)
100%
100%
Automatic I:E
ratio
1:2
1:2
1:2
1:2
1:5
Automatic
frequency
(per minute)
12 to 24
12 to 24
10 to 25
10 to 25
10
Manual flow
rate (L/min)
40
40
40
40
40