4.7 ADDITIONAL CONSIDERATIONS
4.7.1 Low Cylinder Contents. A situation may arise where the Oxygen in the cylinder may reach a
very low level with or without the operator being aware of the fact. The pressure therefore
becomes too low to operate the
microVENT
. The pressure this occurs at is dependent on the
individual pressure regulator characteristics but should not occur above 8 Bar (112.p.s.i.)
indicated cylinder pressure when using a quality regulator.
4.7.2 Cylinder Replacement Pressure. The cylinder replacement pressure is normally indicated on
the regulator pressure gauge. Many gauges indicate the red refill region beginning at 30 Bar
(435 p.s.i.) pressure. It is strongly recommended that once the pressure reaches this section
the cylinder is switched for a fully charged cylinder.
4.7.3 Immersion in liquids. It is recommended that the
microVENT
is not immersed in any liquids at
any time. However accidental immersion may occur. If this event does happen wipe clean the
microVENT
of any detritus picked up from the fluid with a suitable lint free cloth and then allow
the
microVENT
to drain of any liquids. Once all liquids appear to have drained from the body of
the
microVENT
perform functional tests as per section four of this manual.
5 CHAPTER FIVE: SERVICING
CAUTION
The
microVENT
® is designed to provide respiratory support in an emergency situation.
Failure to follow the maintenance and inspection routines properly could result in
incorrect operation of the Microvent.
5.1 ROUTINE MAINTENANCE
5.1.1
To ensure proper operation of the device, regular inspection and checking of the device for
correct function should be undertaken by a responsible member of staff on at least a monthly
basis. This check is to ensure that all Components and Accessories are present, the Oxygen
Cylinder is full and that the
microVENT
is in working order.
5.1.2
The following recommendations for servicing frequency are:
5.1.3
Monthly checks as per the checklist below.
5.1.4
Specification checks and preventive maintenance at 12 months.
5.1.5
In accordance to the Medical Device Alert MDA/2003/007, the Medical Oxygen Supply Hose is
required to be replaced on a 5-yearly basis. If there is any doubt in ascertaining the age of the
oxygen supply hose, please contact B.N.O.S. Meditech Ltd with the oxygen supply hose batch
number as indicated on the hose.
5.1.6
Automatic units (
microVENT
Classic and Responder) should also be checked and run on a
weekly basis
5.1.7
Details of Service Contracts available can be obtained from BNOS Meditech Ltd.