3.4.2 Object Orientation
By the adjustable adaptor (fig. 2, no. 1) at the cassette clamp, which is fixing the object clamps at
the cylinder head, the specimen can be oriented.
figure 2
To adjust the specimen in the desired position, the
clamping lever (fig. 3, no. 2) has to be turned to the
front. Thereby the object clamp is released and can
be turned 360° to find your desired position.
In addition the object clamp can be adjusted by 8°
in each direction (left/right and up/down) by using
the two adjusting screws (fig. 3, no. 1).
When the adjustments are done the clamping lever
(fig. 3, no. 2) has to be pushed back up, so that the
adjusted orientation of the specimen is fixed.
Note: To facilitate the orientation of the specimen,
the clamping lever (fig. 3, no. 2) may be pushed up
slightly. The slight fixing prevents undesired
movements of the clamp during adjustments.
figure 3
To keep the adjusting screws under constant tension, the spring bolt
(fig. 3, no. 3) always has to be tightened.