Maintenance & Cleaning

Control unit

 

WARNING:

 Ensure the Control Unit 

is disconnected from the mains electricity 
supply before cleaning.

 

WARNING:

 DO NOT spray disinfectant 

directly on to the Control Unit, or immerse 
in any type of liquid. This could result in a 
severe electrical hazard as this equipment 
has no protection against ingress of water.

 

WARNING:

 This equipment is not 

suitable for use in the presence of a 
flammable anesthetic mixture with air or 
with oxygen or nitrous oxide.

External cleaning

Wipe down Control Unit with warm water 
containing detergent (or with a solution 
of Sodium Hypochlorite or similar), dry 
thoroughly before use. 

Air Filter Replacement

1.  Switch off the power supply to the 

Control Unit. 

2.  Disconnect the power cord and  

air hoses.

3.  Place the Control Unit on a flat  

surface with its back panel uppermost  
(NOTE: be sure to place soft cloth under 
unit to prevent scratches).

4.  Carefully pry off the cover of the air 

filter, remove and discard the filter 
material and fit new (there may be a 
small locking screw – use a small Philips 
head to remove this first).

5.  Refit the air filter cover to the Control 

Unit. The Control Unit is now ready for 
re-connection.

NOTE: Good filter maintenance is critical to 
maintain your system in optimal operating 
condition. Failure to keep the filters 
clean will result in system downtime and 
increase repair costs. It is recommended 
that the air filter be replaced annually. 
Air filters are available from a Medistore 
Service Centre. Preventative inspection and 
calibration is not required.

Fuse Replacement

1.  Ensure the Control Unit is deactivated.

2.  Remove the power cord from the 

electrical socket on the side of the 
Control Unit.

3.  Using a small flat head screwdriver 

insert into the grove and turn  
anti-clockwise (¼ turn). 

4.  Remove the “blown” fuse from the fuse 

holder clip and discard.

5.  Insert a new fuse into the plug. Push 

against the force of the spring and turn 
clockwise with the screwdriver a ¼ turn.

Содержание ProCair 3000

Страница 1: ...User Guide ProCair 3000...

Страница 2: ...fore operating this medical equipment it is important to read this User Guide and understand the operating instructions and safety precautions Failure to do so could result in patient injury and or da...

Страница 3: ...allation 7 Unpacking Inspection 7 System Installation 8 Control Unit Installation 8 Operation 9 Control Unit Panel 9 Removal and Transport Function 10 Troubleshooting 11 Maintenance and Cleaning 12 13...

Страница 4: ...are isolated from the internal metal motor components with protecting insulation WARNING The double insulated system is intended to protect the user from shock resulting from a break in the Control Un...

Страница 5: ...grounded power receptacle and use the power cord supplied with the system Exposure of the electronic Control Unit to any liquid while it is plugged in could result in a severe electrical hazard Medist...

Страница 6: ...y It is controlled by a simple analogue interface with a dial to control comfort settings and a visible alarm to warn of system low pressure or power failure On the left side of the Control Unit are t...

Страница 7: ...ess Overlay and accessories from the shipping cartons Inspect all items for any damage that may have occurred during shipping Any damage or missing components should be reported to a Medistore Service...

Страница 8: ...lation a Position Control Unit by its hanging hooks over foot board of the bed or on the floor under the bed b Attach two inflation air hoses to Control Unit with the press on quick release connectors...

Страница 9: ...air cells and therefore coming in contact with mattress You should be able to slide your hand between the cells underneath the patient s sacral area bottom along a cell that is deflated Ensure patien...

Страница 10: ...minutes for the setting to take effect 2 Ensure that the overlay is not currently undergoing an alternation If not then turn off Control Unit and disconnect both hoses Immediately connect the air hose...

Страница 11: ...hoses along the inside of the overlay each should be firmly connected Check each air cell is securely attached to the connecting air pipe Check all cells pipes and hoses for any air leakage Check that...

Страница 12: ...TION Do NOT dry the Top Cover using the HEAT cycle or a dryer Air dry or select a low or NON heat dry cycle Washing Alternative If there are no visible signs of body fluids and or substances on the to...

Страница 13: ...to prevent scratches 4 Carefully pry off the cover of the air filter remove and discard the filter material and fit new there may be a small locking screw use a small Philips head to remove this firs...

Страница 14: ...th 2000mm Width 880mm Height 90mm Weight 8 4kgs Control Unit Dimensions Weight Height 115mm Width 135mm Depth 270mm Weight 2 1kgs Power Rating 12VA Cells 21 cells Cell material 0 15mm TPU film laminat...

Страница 15: ...roblem failure is due to customer damage negligence and or misuse unauthorised repairs Items not covered under warranty include but are not limited to stains punctures cuts damages to electrical cords...

Страница 16: ...the CE mark in accordance with EC Directive on Medical Devices 93142lE EC Control Number 95CG Classified by Underwriter s Laboratory Inc for UL 2601 1 standard for Medical Electrical Equipment Medist...

Отзывы: