12

Maintenance & Cleaning

Infection Control and routine cleaning 
must be carried out in accordance with 
your local Infection Control Policy.  
It is suggested that all disinfection  
be done with a high grade  
disinfectant in accordance with  
manufacturer’s instructions.

 

WARNING:

 Do not use high 

temperature autoclave, or use  
Phenolic based products for cleaning. 

IMPORTANT:

 It is recommended the  

system is cleaned between patients  
and approximately every two weeks if  
in constant use.

Overlay Top Cover 
Washing and Disinfecting

If there are visible signs of body fluids  
and/or substances present, top covers 
should be washed. Top covers can be 
machine washed (up to 70°C) using 
disinfectant according to manufacturer’s 
instructions. To determine the amount of 
disinfectant to use, determine the amount 
of water in the washer and then follow the 
manufacturers’ instructions for dilution. 
Soak the Top Cover in the disinfectant 
during the wash cycle. Rinse well in clean 
water and dry thoroughly before use.

 

CAUTION:

 Do NOT dry the Top Cover 

using the HEAT cycle or a dryer. Air dry or 
select a low or NON heat dry cycle.

Washing Alternative

If there are no visible signs of body fluids 
and/or substances on the top cover, it can 
be sanitised.

1.  Apply an intermediate level disinfectant 

(or a solution of Sodium Hypochlorite 
or similar up to 10,000 ppm available 
chlorine) to the Top Cover upper  
surface either by spraying or by  
hand application.

2.  Ensure the surface is completely covered 

with the disinfectant and remains in 
contact with the surface according to 
manufacturer’s instructions.

3.  Remove disinfectant and  

rinse thoroughly.

4.  Allow to air dry before use.

Outside Shell

Wipe down with disinfectant (or a solution 
of Sodium Hypochlorite or similar up to 
10,000 ppm available chlorine), ensuring 
that all surfaces come in contact with the 
disinfectant. Rinse off well with a clean 
damp cloth and air dry.

Air Cell Assembly

 

CAUTION:

 DO NOT machine wash or 

dry the air cells. The air cell assembly does 
not routinely need to be cleaned  
or disinfected.

IMPORTANT:

 DO NOT disassemble the 

overlay, unless cleaning is required.  
If cleaning or disinfecting is required,  
DO NOT disconnect the hoses from the  
air cells.

Содержание ProCair 3000

Страница 1: ...User Guide ProCair 3000...

Страница 2: ...fore operating this medical equipment it is important to read this User Guide and understand the operating instructions and safety precautions Failure to do so could result in patient injury and or da...

Страница 3: ...allation 7 Unpacking Inspection 7 System Installation 8 Control Unit Installation 8 Operation 9 Control Unit Panel 9 Removal and Transport Function 10 Troubleshooting 11 Maintenance and Cleaning 12 13...

Страница 4: ...are isolated from the internal metal motor components with protecting insulation WARNING The double insulated system is intended to protect the user from shock resulting from a break in the Control Un...

Страница 5: ...grounded power receptacle and use the power cord supplied with the system Exposure of the electronic Control Unit to any liquid while it is plugged in could result in a severe electrical hazard Medist...

Страница 6: ...y It is controlled by a simple analogue interface with a dial to control comfort settings and a visible alarm to warn of system low pressure or power failure On the left side of the Control Unit are t...

Страница 7: ...ess Overlay and accessories from the shipping cartons Inspect all items for any damage that may have occurred during shipping Any damage or missing components should be reported to a Medistore Service...

Страница 8: ...lation a Position Control Unit by its hanging hooks over foot board of the bed or on the floor under the bed b Attach two inflation air hoses to Control Unit with the press on quick release connectors...

Страница 9: ...air cells and therefore coming in contact with mattress You should be able to slide your hand between the cells underneath the patient s sacral area bottom along a cell that is deflated Ensure patien...

Страница 10: ...minutes for the setting to take effect 2 Ensure that the overlay is not currently undergoing an alternation If not then turn off Control Unit and disconnect both hoses Immediately connect the air hose...

Страница 11: ...hoses along the inside of the overlay each should be firmly connected Check each air cell is securely attached to the connecting air pipe Check all cells pipes and hoses for any air leakage Check that...

Страница 12: ...TION Do NOT dry the Top Cover using the HEAT cycle or a dryer Air dry or select a low or NON heat dry cycle Washing Alternative If there are no visible signs of body fluids and or substances on the to...

Страница 13: ...to prevent scratches 4 Carefully pry off the cover of the air filter remove and discard the filter material and fit new there may be a small locking screw use a small Philips head to remove this firs...

Страница 14: ...th 2000mm Width 880mm Height 90mm Weight 8 4kgs Control Unit Dimensions Weight Height 115mm Width 135mm Depth 270mm Weight 2 1kgs Power Rating 12VA Cells 21 cells Cell material 0 15mm TPU film laminat...

Страница 15: ...roblem failure is due to customer damage negligence and or misuse unauthorised repairs Items not covered under warranty include but are not limited to stains punctures cuts damages to electrical cords...

Страница 16: ...the CE mark in accordance with EC Directive on Medical Devices 93142lE EC Control Number 95CG Classified by Underwriter s Laboratory Inc for UL 2601 1 standard for Medical Electrical Equipment Medist...

Отзывы: