Medist MEVACS M90 Скачать руководство пользователя страница 9

 

 

 

 

 

 

 
UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009 

 

 

T-QN-04/05-01-1

 

 

9 / 15 

0  /  I 

Switch On-Switch Off 

Control Panel 

 

Turn the right - the vacuum is rising 

Control Panel 

 

Attention, see the User’s Manual 

Control Panel 

 

Foot Control 

Electric Label 

 

Fuse 

Electric Label 

 

5. Unit interconnection 

5.1. Interconnection of the unit with basic equipment 

 

 

Procedure: 
1.  Connect the silicone hose (2) to the hose 

connector VACUUM (1).                                    

2.  Connect the bacteriologic filter (3) to the hose 

(2). The side of filter marked „IN“ was directed 
to the safety jar (6)

3.  Shift the hose (4) onto the other bacteriologic 

filter connector. 

4.  Shift the hose (4) to the output connector (5) 

(safety valve under the connector is safety 
valve (7)) of safety jar. 

5.  Shift the hose (10) on the input connector (9) 

(stream-line tip (8) is under connector) of safety 
jar (6)

6.  Shift the loose end of the hose (10) on the 

output connector (12) of collecting jar - a safety 
valve is under connector - it is labelled as 
“VACUUM”. 

7.  Shift the suction hose (11) on the input 

connector (13) (without safety valve) of the 
collecting jar - it is labelled as “PATIENT”. 

 

 
 

6. Switching on and operation

 

Attention:

 Ensure so as before each suction by new patient was realized sterilization of suction 

hose with suction adapter, vessel for secretion, safety vessel and interconnection hose to 
microbiological filter.
 

6.1. Suction unit operational set-out 

Connect  a  suction  unit  with  a  vessel  exit  connector  to  secret  –  it  is  labelled  as  “VACUUM”.  Hoses  must  not  be 
damaged, cracked or messed up.  

Important!!!  

Check the bacteriological filter from side “IN” and if the filter is glazed replace its.

 

6.2. Vessel preparation for secret 

Vessels must be controlled for possible breaks, granularity or damage before use. Any defect can produce leakage or 
split of the vessel (implosion of the vessel on secret is possible at vacuum). 
Before sucking we recommend to infuse into the secret vessel a little water or disinfecting agent for easier cleaning.

  

10 

11

12 

13 

Содержание MEVACS M90

Страница 1: ...IST s r o Petrusovskeho 4 066 01 Humenne Slovakia T Fax 00421 57 7753276 E mail medist medist sk http www medist sk T QN 04 05 01 1 2070 Medical electric suction unit Ordering number MEVACS M90 72001...

Страница 2: ...service extension are made by MEDIST company or authorized service organizations the device is used in accordance with this instruction manual Use only microbiological filters supplied by manufacture...

Страница 3: ...lectric mains when unplugged from the unit socket Attendant must make certain before every use of suction unit about its good condition and device ready Don t expose the collection vessels for direct...

Страница 4: ...H BASIC EQUIPMENT 9 6 SWITCHING ON AND OPERATION 9 6 1 SUCTION UNIT OPERATIONAL SET OUT 9 6 2 VESSEL PREPARATION FOR SECRET 9 6 3 SWITCHING ON 10 6 4 FUNCTIONAL EXAMINATION OF SUCTION UNIT 10 6 4 1 In...

Страница 5: ...drophobic bacteriological filter 11 Protection of the unit personnel and patient against infection by means of fitted bacteriologic filter 12 The possibility of all functions operation by means of foo...

Страница 6: ...ssure fit lid and metal tubing nipple 42011 12 15 4 0 L jar TM polysulfone with pressure fit lid and metal tubing nipple 42011 10 16 MONOVAC suction liner capacity 1 4 L 42002 01 17 EURO universal rai...

Страница 7: ...x 2 0Lit jar MEVACS M90 2 x 4 0Lit jars 2 x 2 0Lit jars MEVACS M90 2 x 4 0Lit jars 1 x 2 0Lit jar foot switch 3 Installation The MEVACS suction unit is a compact unit There is no needed assembly 3 1 U...

Страница 8: ...els with brake 8 foot control 9 change valve 10 microbiologic hydrophobic filter 11 EURO rail adapter 12 safety jar 13 safety overflow valve 14 hose connector for connection of vacuum branch 15 safety...

Страница 9: ...safety valve is under connector it is labelled as VACUUM 7 Shift the suction hose 11 on the input connector 13 without safety valve of the collecting jar it is labelled as PATIENT 6 Switching on and...

Страница 10: ...acuum 6 4 Functional examination of suction unit 6 4 1 Inner suction circuit turn the suction unit on cover the connector 1 on the suction unit by your finger turn the vacuum regulating vent with the...

Страница 11: ...removed during clearing 6 6 Changing of bacteriologic filter Suction microbiologic filter eliminated risk of contamination with bacteria viruses and infected particle from patient to suction pump or c...

Страница 12: ...2 x 2 0 4 0 LT polycarbonate jars with overflow valve Suction Suction tube Silicone 8 x 3 mm 2 0 m Operation Continuous Filtration Hydrophobic suction microbiological filter with very high bacterial...

Страница 13: ...protect together suction unit for overflow secretion The plastic case holders truck and other equipment can be cleaned with soft wet cloth moistened by cleaning agent which does not react with the pl...

Страница 14: ...ied faults contact either qualified persons of service company or directly the manufacturer MEDIST s r o Petrusovskeho 4 066 01 Humenne Slovakia tel fax 00421 57 7753276 E mail medist medist sk 8 3 Re...

Страница 15: ...2 0 L jar TM with screw metal lid polycarbonate autoclavable 121 O C No 42011 14 polysulfone autoclavable 134 O C No 42011 21 safety valve in lid metal connectors 4 0 L jar TM with pressure fit lid p...

Отзывы: