UP MEVACS M90 - D4O-0085-2-0 / 13.1.2009
T-QN-04/05-01-1
12 / 15
Exchange:
The filter is necessary replaced, as it is glazed with dirtiness or by liquid sucked. Maximum using time is
30 days. After ending of suction of infection patient we recommended immediately exchanged the filter.
It prevents spread of infection.
7. Technical data
Typology MDD/93/43/EEC
Medical Device CLASS IIa
Model
MEVACS M90
Specification by EN ISO 10079-1
High Vacuum - High Flow
Electrical safety
Class I
Direct/indirect contact safety
Type B
Substances penetration protection
IP 20
Safety level in presence of anaesthetic/explosive mixture
Not protected device
Power consumption
220-230V ± 10% / 50 Hz
Input power
420W + 15%
Electric protection
2 fuses Ø 5 x 20 T - 3,15 A
Electric motor
Induction mono-phase with thermic protection
Vacuum pump
Maintenance-free piston vacuum pump
Air flow rate of pump
108 L/min
Air flow rate on socket of suction
unit
89 L/min
Air flow rate on the end of suction
tube in the base version
76 L/min
Maximum vacuum
up to - 92,4 kPa (693mmHg)
Jars
1or 2 x 2,0 / 4,0 LT polycarbonate jars with
overflow valve
Suction
Suction tube
Silicone - Ø 8 x 3 mm - 2,0 m
Operation
Continuous
Filtration
Hydrophobic suction microbiological filter with
very high bacterial efficiencies up to
99,99999% particles bigger than 0,027 micron
Noise level
45 +15% dB (A) @1m
(by ISO 7779)
Weight
33 kg
Dimension (L x W x H)
apr. - 500x500x1080 mm
Supply cord
4 m
Vacuum control valve
Control vacuum gauge (precision ± 2.5%)
Polycarbonate jar ml. graduated - autoclavable
Special features (optional):
-
operation with foot control (IPX8)
Surrounding environment
temperature
5 ÷ 35
o
C
Working condition
Relative humidity
30 ÷ 75%
Surrounding environment
temperature
- 40 ÷ 70
o
C
Storage condition
Relative humidity
10 ÷ 100%
Packing - wrappage
5 laminated cardboard carton
International Standards:
MDD 93/42/EEC; EN 60601-1; EN 60061-1-2; EN ISO 10079-1; EN ISO 14971
Technical life:
10 years
8. Basic maintenance and cleaning
8.1. Unit attending and cleaning
After each use of the MEVACS unit the cleaning and disinfection of secretion vessel and suction hose is
necessary to be performed and you have to perform sterilization. The filter replacement when it has
blocked with liquid or dirty over time. After ending of suction of infection patient we recommended