P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 96
-
For more information on B-mode generic distance measurements, please refer to section
6.6.1
Generic Measurement in B mode.
3.
The result is displayed in the measured result window.
4.
To begin a new GSD-H measurement, repeat steps 1 through 3. You can measure a maximum
of one group of data.
8.4.2. GSD-V
To measure GSD-V:
1.
In the
OB MEAS EQU
menu, roll the trackball to highlight the menu
GSD-V
, press the
Set
key, and move the cursor to image and display “+”.
2.
Measure GSD-V, in the method of distance measurement.
For more information on B-mode generic distance measurements, please refer to section
6.6.1
Generic Measurement in B mode.
3.
The result is displayed in the measured result window.
4.
To begin a new GSD-V measurement, repeat steps 1 through 3. You can measure a maximum
of one group of data.
8.5. Bovine Obstetrics Measurement
When the exam mode
OB
Bovine
is selected, the
OB MEAS BOV
menu will display after
entering the obstetrics measurement.
Figure 8-5 Bovine Obstetrics Measurement and Calculation Menu
After measuring one or more measurement items of CRL, TD, HD, the system will calculate the
MA automatically and display the result.
The measurements of CRL and HD are the same to that of the Canine, please refer to section
and
respectively for reference.
To measure TD:
1.
In the
OB MEAS BOV
menu, roll the trackball to highlight the menu
TD
, press the
Set
key,
and move the cursor to image and display “+”.
2.
Measure TD, in the method of distance measurement.