7
Unplug the device before cleaning
.
Cleaning with water
Wash all nebuliser components (except air hose) under warm tap water (max 60 °C) for about 5 minutes
adding if
necessary a small quantity of detergent following dosage and use limitations as provided by detergent manufacturer. Rinse thoroughly
making sure that all deposits are washed away and leave to dry.
Disinfecting
All nebuliser components (except air hose) can be disinfected with chemical disinfectants following dosage and use limitations as
provided by disinfectant manufacturer. Disinfectants are usually available at pharmacies.
Sterilizing with steam
All nebuliser components (except air hose and masks) can be heat steam sterilized up to 121 °C (20 min.) or 134 °C (7 min.). After
sterilization always let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when components are still warm.
MAINTENANCE, CARE, AND SERVICE
Order all spare parts from your dealer or pharmacist.
It is advisable replace nebulizer after some 100 to 120 treatments on single patient or after about 20 sterilization cycles. Check the filter
continually for cleanliness and replace it if dirty, or after a maximum of 3 months use. Spare filters are provided with the device. To
replace the filter, open the air filter compartment
4
marked
«Filter»
and change the filter
11
.
MALFUNCTIONS AND ACTIONS TO TAKE
The device cannot be switched on
Ensure the power cord
2
is correctly plugged into the socket. Ensure the ON/OFF switch
3
is in the position
«I»
.
The nebuliser functions poorly or not at all
Ensure the air hose
6
is correctly connected at both ends. Ensure the air hose is not squashed, bent, dirty or blocked. If necessary, replace
with a new one. Ensure the nebuliser
5
is correctly assembled
12
. Ensure the required medication has been added.
TECHNICAL DATA
Piston electrocompressor with thermal cut
Nominal voltage: see data label
Dimensions: 130x190x100 mm
Weight: 1.1 Kg approx.
Temporary use: 20/40 min. ON/OFF
MDD 93/42/EEC - Class of risk IIa
Maximum pressure: 185 kPa
Air flow: 10 Lt/min
Noise level: 58 dBA (EN13544-1)
Neb rate: 0.35 ml/min (with 0.9% NaCl solution)
Nebulizer capacity: min. 2 ml - max. 12 ml
Operating pressure: 60 kPa (0.6bar)
Operating flow: 5 l/min
MMAD: 2.61 μm GSD: 2.75 Aerosol Output: 143 μl Output Rate: 87 μl/min
Expected service life: 5 years
Operating Temperature/Humidity: +10° C to +40° C/10% to 95% RH
Storage and Transport Temperature/Humidity/Air Pressure: –25° C to +70° C /10% to 95% RH /690 - 1060hPa
IMPORTANT INFORMATION REGARDING ELECTRO MAGNETIC COMPATIBILITY (EMC)
Electronic devices (PC’s, mobile phones, televisions, etc) that generate electromagnetic interference can affect the correct functioning of
medical devices. This situation can lead to incorrect functioning creating unsafe conditions. In order to show compliance with the Direc-
tives on Electromagnetic Compatibility (EMC), all the tests introduced by the standard EN60601-1-2:2015 have been implemented. This
standard defines the levels of immunity to electromagnetic interference as well as maximum levels of electromagnetic emissions for me-
dical devices:
3A HEALTH CARE
products comply with the levels specified in the standard. However, when the device is in use, it is best
practise not to use mobile phones or other electronic devices which generate strong electromagnetic fields near the medical device. This
may create dangerous conditions that put the successful outcome of treatment at risk. On the basis of experience and tests conducted, it
is recommended that a minimum distance of 7 m be kept: if the distance is less, the correct functioning of the device should be verified.
Documentation regarding the standard EN60601-1-2:2015 is available from
3A HEALTH CARE
(see the address in the instruction
manual) and on the website
www.3-a.it
.