Rev. D 2018-11
26
3. Diagnosis
An extensive autodiagnosis system has been implemented into the oxygen concentrator
Précise 6000
. By pressing the button next to the oxygen outlet on the right the diagnosis
is called up. The displays can be called up by pressing the button.
Servicebutton
press x times
Designation
Information
1
Language ENGLISH
Select language
2
Softwareversion
<<V1.56 –M->>
Software version & equipment
e.g. Précise 6000M
3
Serial number
Individual serial number on the
equipment
4
O2 Sensor value
Temperature and
concentration value
5
Flow Sensor value
Flow value l/min
6
Power switch inverts
Factory adjusted
7
Concentration monitoring:
< 80% O2
Factory adjusted
8
Concentration monitoring:
< 72% O2
Factory adjusted
9
Battery monitoring
Factory adjusted
10
O2 < 82% Ionization disconnection
Factory adjusted
11
O2 Concentration display
Factory adjusted
12
Flow to small (Monitoring)
Factory adjusted
13
Flow to high (Monitoring)
Factory adjusted
14
Equipment version
Factory adjusted
15
Service telephon number
Servicetelefonnummer des
Fachhandelspartners
16
Interval time
Factory adjusted
17
Date
Adjusted
18
Time
Adjusted
19
Working hours counter
1 = Patient 2 = Total
20
min. of pressure values of all 3
columns
Defect of main valve, columns,
sensor, compressor etc.
21
max. of pressure values of all 3
columns
Defect of main valve, columns,
sensor, compressor etc.
22
Error memory bank of last 10
Alarms
Indicates the error with date
and time