EasyOne Pro™ / EasyOne Pro™ LAB
51
Performing a Test
FVL Test (inspiratory and expiratory measurement)
Ambient Conditions
The ATPS values (
A
mbient
T
emperature
P
ressure
S
aturated = spirometer
conditions) of inspiratory measurements must be converted to BTPS values
(
B
ody
T
emperature
P
ressure
S
aturated = body conditions). For this
conversion, the system needs the following data:
relative humidity
room temperature
atmospheric pressure or altitude.
Before the inspiratory test can be started, a window pops up with these
data. Check the data. For the pending test you can now edit the data, if
necessary, or confirm them, if they are correct.
Edit the values, if necessary, and click
Confirm >>
to close the window.
Warning
Erroneous Measurement Results—Incorrect entries may lead
to erroneous test readings and incorrect system
interpretations. The specified measuring accuracy is ensured
only when all entries are correct.
Fig. 3-14
Window showing the ambient conditions
a b c
d
e
f
a Conversion factors
b Box for entry of the humidity
c Box for entry of the ambient temperature
d Box for entry of the altitude (atmospheric
pressure)
e Click to recalculate the BTPS factor (if the
ambient conditions were modified)
f Click to confirm and accept the values