Medical Freezer MPF1.8L 80 Скачать руководство пользователя страница 8

- 6 -

Before using the freezer for the first time, the moisture in the device should be cleaned, especially the water on
the inner door and inner door cover to prevent the door from freezing when the temperature is low, which
makes the door difficult to open!

Any product may be damage. When you store more important items, be sure to take appropriate temperature
monitoring method. In order to further monitor the operation status of Meiling series scientific research and
medical freezers, our company has developed more optional functions such as the RS485 remote monitoring,
wireless monitoring, and the short message alarm function and etc..

In order to achieve rapid cooling and stable storage of the sample, please use the cold storage block attached
to the product. After removing the cold storage block, the product can still work normally, but the product
performance will be reduced.

Note: The inner cover of the ultra-low temperature freezer can be effectively insulated. When
using the ultra-low temperature freezer, please cover the inner cover of the device!

The ultra-low temperature freezer is suitable for low temperature test of special materials, frozen red blood
cells, white blood cells, skin, bones, bacteria, biological products, ocean products, electronic devices, etc.,
suitable for hospitals, epidemic prevention stations, blood banks, universities, bio-pharmaceuticals, genetic
engineering, chemical industry, military enterprises, research institutes, enterprises, etc.. The storage
temperature in the cabinet can be adjusted through the control buttons on the display, which is convenient to
use and reliable in performance.

V. Preparation before Use

Please read this manual carefully for the first time using!

Содержание MPF1.8L 80

Страница 1: ...OPERATION MANUAL Applicable product model MPF1 8L 80 Medical Freezer Ltd Info medical freezer...

Страница 2: ...s 7 1 Display Panel Function Descriptions 7 2 First Initial Starting 8 3 Function Setting 8 VII Defrosting Unused and Maintenance 10 1 Defrost 10 2 Stop Using 10 3 Maintenance 10 VIII After sales Serv...

Страница 3: ...Freezer Only NANBEI or agents authorized by NANBEI could maintain this device If the operator encounters a situation not mentioned in this manual please contact the authorized agent of NANBBEI Biomed...

Страница 4: ...2...

Страница 5: ...3...

Страница 6: ...4...

Страница 7: ...s Be extremely careful not to use volatile solvents such as gasoline to wipe plastic rubber parts After running for a period of time in the ultra low temperature freezer a layer of frost will form on...

Страница 8: ...the product After removing the cold storage block the product can still work normally but the product performance will be reduced Note The inner cover of the ultra low temperature freezer can be effec...

Страница 9: ...damage the system Normal working conditions a Ambient temperature 10 C 32 C b Relative humidity 80 c No strong vibration and no corrosive gas around d No direct sunlight or any other cooling or heati...

Страница 10: ...will appear on the screen The initial password is 000000 iii Click the switch button to operate the device 3 Function Setting Regular display In normal use the current temperature is displayed as the...

Страница 11: ...word is entered the left curve number will be displayed The alarm code information is as follows Do not put the items in the ultra low temperature freezer that has just been connected to the power sup...

Страница 12: ...oth to dry the water around the cabinet especially the part where the door seal is in direct contact with the cabinet to avoid freezing the door seal at low temperature 3 Maintenance Clean and maintai...

Страница 13: ...e otherwise the hot air enters the inside of device and the negative pressure caused by the sharp drop in temperature causes the door to be difficult to open Once there is abnormal alarm when turn on...

Страница 14: ...ators 1 Main Technical Parameters Model MPF 1 8L 80 Temp range 40 C to 80 C Volume 1 8 L Style vertical with handle External dimensions width depth height 245 282 496mm Internal dimensions width depth...

Страница 15: ...rrent A 0 46A Power 80W 2 Packing List Name Operation manuel Quantity 1 3 Electrical Schematic Diagram Due to the rapid improvement of the product please refer to the technical data and electrical sch...