Medical Freezer MPF1.8L 80 Скачать руководство пользователя страница 11

- 9 -

000000. After inputting, you can change the password. Click the “ON” or “OFF” next to the password lock to
enable or disable the password lock.
iii. Alarm function: In the power-on state, the device will alarm when there is any of the following conditions,
there will be a beeping alarm: 1 the ambient temperature is higher than 40 ° C or lower than 0 ° C; 2 hot end
temperature is higher than 55 ° C; 3 the current temperature differs from the target temperature by more than
5 °C; At this point, you can press the mute button to eliminate, please pay attention to the alarm information
and deal with it in time.
iv. Exit the setup status: Click the Back button in the setting state to return to the main page and exit the setting
state.
v. After the temperature setting is completed, click the power button to run.

Remark: In the locked state, after entering the password input interface without losing the password or the
password input error returns to the main page, the left curve number will be hidden; when the unlock input
correct password is entered, the left curve number will be displayed.

The alarm code information is as follows:

Do not put the items in the ultra-low temperature freezer that has just been connected to the power
supply, let the empty box run for a while (about 5 hours), then store the frozen items in the device.

Each item cannot exceed 1/3 of the device capacity. After reaching the set temperature and running
for 5 hours, put another 1/3 of the device capacity items for storage.

Try not to open the door during the cooling process, otherwise it will cause the temperature to rise.

If there is a device failure or power failure, the temperature in the low temperature device will rise. If
it cannot be repaired in a short time, please take out the stored items and transfer them to other
places that meet the temperature requirements of the stored items to avoid damage.
Due to the cooling inertia, the display temperature and the set temperature may be different when
the product reaches the temperature set value (especially when the temperature setting is low),
which is normal.
Do not set the temperature frequently in a short time, otherwise the inertia of hot and cold may be
too large, and may not achieve your desired adjustment effect.

Содержание MPF1.8L 80

Страница 1: ...OPERATION MANUAL Applicable product model MPF1 8L 80 Medical Freezer Ltd Info medical freezer...

Страница 2: ...s 7 1 Display Panel Function Descriptions 7 2 First Initial Starting 8 3 Function Setting 8 VII Defrosting Unused and Maintenance 10 1 Defrost 10 2 Stop Using 10 3 Maintenance 10 VIII After sales Serv...

Страница 3: ...Freezer Only NANBEI or agents authorized by NANBEI could maintain this device If the operator encounters a situation not mentioned in this manual please contact the authorized agent of NANBBEI Biomed...

Страница 4: ...2...

Страница 5: ...3...

Страница 6: ...4...

Страница 7: ...s Be extremely careful not to use volatile solvents such as gasoline to wipe plastic rubber parts After running for a period of time in the ultra low temperature freezer a layer of frost will form on...

Страница 8: ...the product After removing the cold storage block the product can still work normally but the product performance will be reduced Note The inner cover of the ultra low temperature freezer can be effec...

Страница 9: ...damage the system Normal working conditions a Ambient temperature 10 C 32 C b Relative humidity 80 c No strong vibration and no corrosive gas around d No direct sunlight or any other cooling or heati...

Страница 10: ...will appear on the screen The initial password is 000000 iii Click the switch button to operate the device 3 Function Setting Regular display In normal use the current temperature is displayed as the...

Страница 11: ...word is entered the left curve number will be displayed The alarm code information is as follows Do not put the items in the ultra low temperature freezer that has just been connected to the power sup...

Страница 12: ...oth to dry the water around the cabinet especially the part where the door seal is in direct contact with the cabinet to avoid freezing the door seal at low temperature 3 Maintenance Clean and maintai...

Страница 13: ...e otherwise the hot air enters the inside of device and the negative pressure caused by the sharp drop in temperature causes the door to be difficult to open Once there is abnormal alarm when turn on...

Страница 14: ...ators 1 Main Technical Parameters Model MPF 1 8L 80 Temp range 40 C to 80 C Volume 1 8 L Style vertical with handle External dimensions width depth height 245 282 496mm Internal dimensions width depth...

Страница 15: ...rrent A 0 46A Power 80W 2 Packing List Name Operation manuel Quantity 1 3 Electrical Schematic Diagram Due to the rapid improvement of the product please refer to the technical data and electrical sch...

Отзывы: