Periodically check the condition of the filter. The filter was inserted to protect the
compressor. Correct filter maintenance prolongs the life of the nebulizer. The frequency
with which filter should be replaced depends on the conditions in which the unit is used.
The filter is located on the front of the nebulizer.
To replace the filter:
1. Open the filter cover cap;
2. Remove the filter to be replaced from its housing;
3. Carefully fit the new filter;
4. Fit the filter cover cap back on.
Warning!
Never leave the filter cover cap and/or filters unattended: These are small parts
posing the risk of suffocation if swallowed by children.
Checking and replacing filters
Nominal voltage:
230 V
Frequency:
50 Hz
Power:
< 180 VA
Fuse:
F 5 AT 250 V
Maximum flow:
17 ± 2 l/min
Operational flow:
8 ~ 9 l/min
Max pressure:
> 29 Psi
Operational pressure:
10 ~ 16 Psi
Protection against liquid seepage:
IP21
Weight:
1,66 kg
Dimension (L x W x H):
237,6 × 189,8 × 129,6 mm
Conditions for using the device:
30' ON – 30' OFF
MMAD: – aerosol size:
2,3 μm
PORR (Breathable fraction;
aerosol % smaller then 5 μm):
90,2 % ± 1,2 %
Nebuliser cup capacity
MAX 8 ml
Noise level at 1 m:
< 60 dB
Nebulization rate:
0,2 ~ 0,3 ml/min
Medicine leftover:
< 1 ml
Medical device classification:
IIb
Technical specifications
EN
14