Cleaning and maintenance
Cautions:
Check the medication cup and air filter before each use. Dirty or worn parts should be
replaced. Do not immerse the unit in water.
It is important to clean your Ultrasonic Nebulizer before it is used for the first time and after
each use. Remember to disconnect the unit from the electrical outlet. Never immerse the
unit in water. It may damage the unit.
Do not use any powdered cleaners or soap pads, which may damage the finish.
Cleaning the device
1. Disconnect the unit from the electrical outlet before cleaning.
2. The surface of the main unit should be cleaned with a soft cloth moistened with a
non-abrasive cleanser and water below 30°C. Don’t put the main unit directly into water.
The ultrasonic part can be wiped with a cotton stick with clean water below 30°C.
3. Clean the atomizing nozzle with hot water after each use and clean with a mild
detergent after the last treatment of the day.
Caution! During cleaning procedure, do not press on the ultrasonic membrane,
because it can be permanently damaged.
Filter change
1. Do not use cotton or any other materials. Do not wash or clean the filter. Only use
filters supplied by your distributor. And do not operate without a filter.
2. Change the filter every 30 days or when it turns grey.
3. Changing procedure:
a) Remove the filter cover
b) Replace the used filter with a new one
c) Replace the filter cover
How to preserve the effectiveness of your device
1. Your M480 ultrasound system for aerosol therapy does not require any particular
maintenance care; you only need to:
- avoid hitting or dropping it;
- avoid opening it under any circumstances;
-
keep it in a suitable place, avoiding high temperatures, direct sunlight, humidity and dust;
- avoid using the device for more than 30 minutes uninterruptedly.
2. After all sessions, carefully clean all the accessories used.
Accessories may be washed under running water. If you want to sterilize them, a
cold solution suggested by your family doctor may be used.
3. We recommend performing a maximum of 2 treatments using a single medicinal
holder (cup) before replacing it.
4. We strongly recommend replacing the air filter after 15/20 treatments, or when the
filter becomes grey. For the substitution, proceed as follows:
- remove the cap from its slot under the base of the device;
- remove the old filter and introduce the new one;
- replace the cap inside its slot.