40
Guidance and manufacturer’s declaration – electromagnetic emissions
This breast pump is intended for use in the electromagnetic environment specified below. The
customer or the user of the breast pump should assure that it is used in such an environment.
Emission tests
Compliance
Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The breast pump uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The breast pump is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Pst < 1.0
15. EMC / Technical Description (cont.)
Warning
– This breast pump should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, this breast pump should be
observed to verify normal operation in the configuration in which it will be used.