13
EN
In this table you will see the recommended hours for a tube change on the left.
When a tube change is performed, the actual hours on the hour meter (“total h“)
should be filled in the appropriate column. Also included are two columns for
inserting date and signature.
• Unplug the light unit from the power supply.
• Remove the baby support from the light unit.
• Release the 10 screws with a screwdriver and remove the plexiglass cover.
• Lift the metal clamp and carefully remove the light tube from its socket and
from its clip.
• Put the tube aside immediately in order to avoid confusion.
• Insert the new tube into the clip and push into the socket until it snaps in.
• Readjust metal clamp for a secure hold.
Warning
Replacement tubes must be of the type used by Medela. Otherwise the safety and
success of the therapy cannot be guaranteed! We recommend changing the light
tube every 1500 hours in order to guarantee therapeutic effectiveness.
• Check that the reflector is clean and shiny, correct if not.
• Re-mount the plexiglass cover with the 10 screws.
10.2 Cleaning and disinfecting
• Unplug the unit from the power supply.
Light unit and baby support
The outside surfaces can be cleaned with a damp cloth. For disinfecting, alcohol may
also be used. Cleaning or disinfecting agents should not be used in concentrated form!
If the baby support shows signs of wear, sags or is damaged, replace with a new one.
Important
Do not use cleaning solutions such as appliance disinfectants which use formaldehyde.
Do not autoclave the baby support or disinfect by using heat.
Bilicombi therapy blanket
If you are using a Disposable Bilicombi, discard the Bilicombi after 24 hours or if it
becomes soiled. The Disposable Bilicombi is designed for single patient use.
If you are using a terry cloth, Washable Bilicombi, follow the cleaning instructions below:
• Remove Bilicombi from BiliBed.
• Zipper Bilicombi and attach under-chin Velcro fastener.
• Wash for Standard Machine cycle. Minimum wash cycle of 10 min is recommended.
• Machine wash in hot water. We recommend a temperature of 140 degrees F (60 °C).
The maximum temperature is 210 degrees F (95 °C).
• Standard cleaning detergent to be used. Refer to washing machine
manufacturer for instructions.
• Use of Chlorine Bleach can cause color to bleed and/or fade and can reduce life
of Bilicombi.
• Tumble dry low heat.
• If any of the Velcro fasteners shows decreased adherence or the light-permeable
fabric is damaged/too stiff a new Bilicombi should be used.
10.3 Troubleshooting
If the light goes out during therapy:
• Check if the fan is working properly. If it does not run at the proper speed or if it
is blocked, an electronic safety circuit switches the lamp off in order to protect
the system from overheating. Replace fan if necessary.
• If necessary, change tube.