Electromagnetic Immunity
The Medela Apgar Timer is intended for use in the electromagnetic environment specified below.
The customer or the user of the Medela Apgar Timer should assure that they are used in such
environment.
Immunity Tests
Electrostatic
Discharge (ESD)
IEC 61000-4-2
IEC 60601
test level
± 6 kV contact
± 8 kV air
Compliance
level
± 6 kV contact
± 8 kV air
Electromagnetic
environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humi-
dity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV
for power
supply lines
± 1 kV
for input /
output lines
± 2 kV
for power
supply lines
± 1 kV
for input /
output lines
not applicable
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV
differential mode
± 2 kV
common mode
± 1 kV
differential mode
± 2 kV
common mode
not applicable
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips,
short interrup-
tions and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
not applicable
Mains power quality should be that
of a typical commercial or hospital
environment.
If the user of the electric breast
pump Freestyle requires continued
operation during power mains in-
terruptions, it is recommended that
the electric breast pump Freestyle
is powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Electromagnetic compatibility (EMC, IEC 60601-1-2:2007 3rd Edition, Table 2)