![MEDCHECK SFBG01 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/medcheck/sfbg01/sfbg01_user-manual_1759419021.webp)
When symbol flashes, press
S
button to activate
Quality Control Mode.
With symbol on, your meter is ready to perform
quality control test.
Quality Control of the System
21
3
5
4
Shake the vial of Glucose Control Solution well before
use. Squeeze the vial but discard the first droplet.
Squeeze the second droplet to your clean fingertip or
on a clean object such as a transparency.
Move the collection zone of test strip close to Glucose
Control Solution droplet.
Collection zone automatically drink the droplet in.
When full, your meter beeps and counts down for 5
seconds.
When results match with acceptable range printed on
the Test Strip vial, your System is working properly.
Note:
Check the Test Strip vial for acceptable ranges
for High and Low Levels. Refer to Glucose
Control Solution instruction before use.
Note
To avoid contamination,
do not apply the control
solution to the test strip
directly from the bottle.
Understanding Your Test Results
When results match with acceptable range printed on the Test Strip vial, your
System is working properly, and your test results are reliable.
If your results do not match with the acceptable range, please check the following:
Are the control solutions or test strips within the expiration date?
Has the control solution been contaminated?
Is the code number correct?
Did you follow the test procedures correctly?
Please repeat the test. If results still do not match, contact Smartfuture Pte
Ltd or the authorized distributor.