iNstallatiON
4.2 Device Connection
the device shall be installed on a dental chair. We recommend to apply the wiring scheme as
shown in Figure 4 at page 9
, that is featured with 24 v~ or 32 v supplied by a power supplier
with double insulation compliant to IEC/EN 60601-1, controlled by the switch on the quiver and
protecting the line by a fuse.
fuse
t 315 mal
250 V
tool-holder
starlight s + Cord
Clamps
Black
red
White
Fusebox, fuse, tool-holder
and clamps are not provided
by Mectron
Power supply
24 V~ 50/60 hz
o 32 V
tearing-resistant
line
Figure 4 –
Device connection
NOTE:
the device is not equipped with a disconnection device from the power network
according to clause 8.11.1 of IEC/EN 60601-1, as it is a component that will be installed on a
dental chair in its final configuration. Such a device must be designed for the final installation
and the manufacturer of the dental chair shall take care of assuring the compliance to this
requirement.