Mains Power Supply
Vortex
-dV
test stands can be used on 110–120 or 220–240 Vac 50-60 Hz supplies. The rear
fuse carrier is set for your local requirement but is reversible, so should you replace a
fuse, the correct local voltage must be selected.
The voltage that is selected is indicated by which the arrow is pointing to the white line
located at the bottom of the device. This is illustrated in the image below, shown within
the red circle:
Fuse Specification
A Vortex
-dV
test stand uses two 2A - Speed T, 5 x 20mm fuses. If replacing a blown fuse
only replace the fuse on the active side of the inlet filter with the fuse specified above,
or equivalent.
If you are in doubt, please contact your local Mecmesin support agent for more
information.
Removing the fuse holder