3.3.2
3.3.3
are displayed on the CFG+ and recognised by the symbol (See Fig. 3).
Zeroing the gauge
During the operation of the gauge it is often necessary to zero the display - e.g. when you wish to tare out the
weight of an accessory that you do not want to become part of the measured reading. Press and release the
ZERO
ZERO
key.
When the gauge is in the process of zeroing itself a row of segments on the top and bottom of the display will
be seen to alternately blink once. When this is completed the display will read zero. This function will not be
performed if the gauge is in an overloaded state.
Changing the unit of measurement
You can choose from the following units of measurement depending on the capacity of the gauge: newtons (N),
kilogram-force (kg), millinewtons (mN), gram-force (g), ounce-force (oz), pound-force (lb) or kilonewtons (kN).
To change the display units press and release the
UNITS
UNITS
key. Each successive key press will select the next
available units until the gauge returns to its original setting.
The CFG+ automatically converts readings as new units of measurement are selected.