If the % offset is between 5% - 10% please contact your
supplier to arrange a recalibration of your BFG-HS.
If the % offset is greater than 10% please contact your
supplier to arrange for a loadcell replacement. These values
are given as an indicator only – the need for
calibration/repair may vary according to the individual
characteristics of the load cell.
MAX
-2.7:%
0.5:%
Fig. 4
upper value = offset
when last calibrated
lower value = offset
currently present on
gauge