MEATEST
M-140 Multifunction Calibrator
Operation manual v46
117
If Option 140-41 cable adapter is used for simultaneous calibration and OUTPUT 140-41 option is set to
PANEL, the following generation/measurement ranges can be set:
Source
DC/AC V
0 – 1000 V
DC/ AC I
0 – 20 A
Resistance
0 – 50 M
RTD sensor simulation
Pt 1.385, Pt 1.392, Ni
TC sensor simulation
K, N, R, S, B, J, T, E
Frequency
0.1 – 20 MHz
Power/Energy
0.2 – 240 V / 0.2 – 20 A
Meter
DCV
0 – 12 V
MVDC
0 – 2 V
DCI
0 – 25 mA
Resistance
0 – 2 k
Frequency
1 – 15 kHz
When 140-41 cable adapter is connected, only harmonic signals are generated (SHAPE function is inactive).
Use of Option 40 and Option 60 cable adapter
Opt. 40 cable supports simultaneous mode - measurement of voltage up to 12 V, current up to 25 mA and
frequency up to 15 kHz. Opt. 60 cable is designed for four-wire measurement of resistance or temperature using
resistance temperature sensors. The generation ranges are the same as when 140-41 cable adapter is used.
Use of Option 70
Adapter is aimed for four-wire generation of resistance values and for four-wire resistance temperature sensor
simulation. Re-calibration of resistance function of the calibrator can be performed only with this adapter.
Resistance ranges are the same as two-wire resistance ranges with output on Hi – Lo terminals on front panel.
When adapter Option 70 is used, resistance output is only on output terminals on this adapter. Front panel Hi –
Lo terminals are switched off.
When adapter Option 70 is used, set resistance value is connected to the adapter output terminals permanently.
Status of button ON/OFF doesn’t influence resistance connecting or disconnecting.
Examples of tests
Examples of applications
Generation
Measurement
Application
Pt 100
10 V / 20 mA / f
Calibration of temperature evaluation units, setting of
temperature regulators
TC
10 V / 20 mA / f
Calibration of temperature evaluation units, setting of
temperature regulators
Frequency
10 V / 20 mA
Setting/calibration of energy meters
Resistance
10 V / 20 mA
Measurement of resistance bridges
1. Calibration of industrial thermometers with Pt100 sensor and 20 mA / 10 V output: