M-140 Multifunction Calibrator
MEATEST
98
Operation manual v46
MEASure
?
: CONFigure
: VOLTage
: CURRent
: MVOLTage
: RESistance
: FREQuency
: TEMPerature
: RTD
: TYPE (?)
<CPD> { PT385 | PT392 }
: NRESistance (?)
<DNPD>
: THERmocouple
: TYPE (?)
<CPD> { B | E | J | K | N | R | S | T }
: RJUNction (?)
<DNPD>
: SGS
: VOLTage (?)
<DNPD>
: OFF
MEAS ?
This command returns the measured value.
Example:20.5
is returned as 2.001.
MEAS :CONF ?
M140 returns the set measurement mode { VOLT | CURR | MVOLT | RES | FREQ | TEMPerature:RTD |
TEMPerature:THERmocouple | SGS | OFF }.
MEAS :CONF :VOLT
This command selects the function VOLT (range 0 .. 12V) of internal multimeter and switches it on.
Example: to set the multimeter to DC voltage 10V measurement mode:
MEAS :CONF :VOLT <cr>
MEAS :CONF :CURR
This command selects the function CURR (range 0 .. 25mA) of internal multimeter and switches it on.
Example: to set the multimeter to DC current 20mA measurement mode:
MEAS :CONF :CURR <cr>
MEAS :CONF :MVOLT
This command selects the function MVOLT (range 0 .. 2000mV) of internal multimeter and switches it on.
Example: to set the multimeter to DC voltage 2000mV measurement mode:
MEAS :CONF :MVOLT <cr>
Содержание M-140