![Meatest 9010+ Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/meatest/9010/9010_user-manual_1758451028.webp)
High resistance decade offers fully programmable series connected resistance with maximum working
voltage 1500 V. The decade is applicable on DC voltage. Do not use AC test signal. High resistance
decade is available only in two-wire configuration.
Push MODE soft key to display list of modes. Select High resistance item and confirm with ENTER.
1.
Connect the ohmmeter (multimeter) to the calibrator as shown on the figure below.
2.
Select appropriate function and range on the ohmmeter (multimeter)
3.
Push RESISTANCE function key on the calibrator. Push MODE soft key and select position
High resistance decade.
4.
Set requested resistance value.
5.
Switch output terminals on by pushing key OPER.
6.
Push STBY key to switch output off.
Resistance SETUP can be activated using SETUP SOURCE key. Following parameters are available:
-
Terminal ground On/Off
LO output terminals is internally connected to PE wire when parameter is set to ON.
Calibrator contains continuously adjustable decade, based on capacitance simulation using electronic
circuits. The function is designed for calibration of standard resistance function of various multimeters
which use low level signals. The mode is signed VARIABLE.
Except the VARIABLE mode the calibrator offers more accurate resistance in FIXED mode. In this
mode only fix nominal decadic values are available.
Terminals
HI
LO
I+
I-
HIR
GND
Entry
Function
TRG
HF
9010
Multifunction Calibrator
V
COM
A
R
METER
SCOPE
SOURCE
METER
NULL
SELECT
SETUP
SYSTEM
1100V
RMS
MAX
1500V
DC
MAX
1V PK
MAX
20V PK
MAX
1 00V PK
MAX
10V PK
MAX
2V PK
MAX
25V PK
MAX
10V PK
MAX
50V PK MAX
50 V PK MAX
FUNC
OPER
STBY
10V
RMS
MAX
Содержание 9010+
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 12: ...See 9010 SCPI manual for complete SCPI reference more details on communication setup and troubleshooting ...
Страница 33: ...Following items can be modified ...
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 55: ......
Страница 57: ......
Страница 58: ......
Страница 78: ...Temperature range 250 00 1820 00 C Cold junction comp modes Manual ...