9010+ Multifunction calibrator can be ordered with integrated options as listed in Table 1. Integrated
options can be fitted only by manufacturer and certified service providers.
The calibrator can be integrated into automated calibration systems (ATS) and generally controlled from
remote computer via following remote control interfaces:
-
RS232
-
USB
-
GPIB (IEEE488)
-
LAN
When controlled remotely, maximum ratings of calibrator’s output signals as well as all other
specifications are the same as in manual mode.
Meatest software package Wi Caliber is recommended for best automation results. This system
is designed for automated and semi automated calibrations of digital and analogue meters including
uncertainty calculation, result evaluation and certificate printing according to ISO 17025 standard.
Only one interface can be used for communication at any given time. Default active interface is RS232,
other interfaces can be selected in MENU->Interface->Active bus. To establish connection between the
calibrator and computer, set interface settings in your computer accordingly:
-
COM port
see available COM ports in Windows Device Manager
-
Baudrate USB
according to MENU > Interface > USB Baudrate (9600 by default)
-
Baudrate RS232
according to MENU->Interface->RS232 Baudrate (9600 by default)
-
Data bits
8
-
Stop bits
1
-
Parity
None
-
Handshake (XON/XOFF) Off
-
GPIB Address
according to MENU->Interface->GPIB Address (2 by default)
DHCP service discovery is active by default and is recommended for easiest connection. If you prefer
to use fixed IP address or change any other default settings, go to MENU->Interface->LAN Settings.
LAN communication uses Telnet protocol. Default log-in for Telnet client is “9010+_SNxxxxxx 23”,
where “xxxxxx” stands for serial number and 23 is default communication port.
Содержание 9010+
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 5: ......
Страница 12: ...See 9010 SCPI manual for complete SCPI reference more details on communication setup and troubleshooting ...
Страница 33: ...Following items can be modified ...
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 55: ......
Страница 57: ......
Страница 58: ......
Страница 78: ...Temperature range 250 00 1820 00 C Cold junction comp modes Manual ...