USB-2416 User's Guide
Specifications
26
Table 7. R type thermocouple accuracy specifications, including CJC measurement error.
All specifications are (±).
A/D data rate Sensor temperature
range
Accuracy error,
maximum
Accuracy error,
typical
Tempco
(°C/°C)
3750 S/s
-50 °C
4.826 °C
3.133 °C
0.082
250 °C
2.117 °C
1.424 °C
1768 °C
2.842 °C
2.347 °C
2000 S/s
-50 °C
4.826 °C
3.133 °C
0.082
250 °C
2.117 °C
1.424 °C
1768 °C
2.842 °C
2.347 °C
1000 S/s
-50 °C
4.826 °C
3.133 °C
0.082
250 °C
2.117 °C
1.424 °C
1768 °C
2.842 °C
2.347 °C
500 S/s
-50 °C
4.065 °C
2.379 °C
0.082
250 °C
1.805 °C
1.113 °C
1768 °C
2.619 °C
2.123 °C
100 S/s
-50 °C
4.065 °C
2.379 °C
0.082
250 °C
1.805 °C
1.113 °C
1768 °C
2.619 °C
2.123 °C
60 S/s
-50 °C
4.065 °C
2.379 °C
0.082
250 °C
1.805 °C
1.113 °C
1768 °C
2.619 °C
2.123 °C
50 S/s
-50 °C
4.065 °C
2.379 °C
0.082
250 °C
1.805 °C
1.113 °C
1768 °C
2.619 °C
2.123 °C
25 S/s
-50 °C
4.065 °C
2.379 °C
0.082
250 °C
1.805 °C
1.113 °C
1768 °C
2.619 °C
2.123 °C
10 S/s
-50 °C
4.065 °C
2.379 °C
0.082
250 °C
1.805 °C
1.113 °C
1768 °C
2.619 °C
2.123 °C
5 S/s
-50 °C
4.065 °C
2.379 °C
0.082
250 °C
1.805 °C
1.113 °C
1768 °C
2.619 °C
2.123 °C
2.5 S/s
-50 °C
4.065 °C
2.379 °C
0.082
250 °C
1.805 °C
1.113 °C
1768 °C
2.619 °C
2.123 °C
To achieve the thermocouple accuracies listed above, the USB-2416 should be warmed up for 45 minutes after
the initial power on. The accuracies listed above are only guaranteed if the USB-2416 is housed in its plastic
enclosure.