PCI-DAS6031 & PCI-DAS6033 User's Guide
Specifications
Accuracy
100 kS/s sampling rate, single channel operation and a 15-minute warm-up. Accuracies listed are for
measurements made following an internal calibration. They are valid for operational temperatures within ±1°C
of internal calibration temperature and ±10°C of factory calibration temperature.
Calibrator test source high side
tied to Channel 0 High and low side tied to Channel 0 Low. Low-level ground is tied to Channel 0 Low at the
user connector.
Table 2. Absolute Accuracy
Table 2
Range Absolute
Accuracy
±10 V
±3.76 LSB
±5 V
±13.61 LSB
±2 V
±13.70 LSB
±1 V
±13.83 LSB
±500 mV
±14.09 LSB
±200 mV
±16.71 LSB
±100 mV
±19.99 LSB
0 to 10 V
±6.40 LSB
0 to 5 V
±26.11 LSB
0 to 2 V
±26.28 LSB
0 to 1 V
±26.54 LSB
0 to 500 mV
±27.13 LSB
0 to 200 mV
±32.11 LSB
0 to 100 mV
±38.67 LSB
Table 3. Absolute Accuracy Components – All values are (±)
Noise +Quantization (µV)
Range %
of
Reading
Offset
(µV)
Single Pt
Averaged
1
Temp
Drift
(%/ºC)
Absolute
Accuracy at
FS (mV)
±10 V
0.0061
479.2
634.1
54.9
0.0001
1.147
±5 V
0.0361
243.6
317.1
27.5
0.0006
2.077
±2 V
0.0361
102.2
126.8
11.0
0.0006
0.836
±1 V
0.0361
55.1
63.4
5.5
0.0006
0.422
±500 mV
0.0361
31.6
36.8
3.2
0.0006
0.215
±200 mV
0.0411
17.4
22.5
2.0
0.0006
0.102
±100 mV
0.0461
12.7
19.6
1.8
0.0006
0.061
0 to 10 V
0.0061
326.6
417.8
36.6
0.0001
0.976
0 to 5 V
0.0361
167.3
208.9
18.3
0.0006
1.992
0 to 2 V
0.0361
71.7
83.6
7.3
0.0006
0.802
0 to 1 V
0.0361
39.9
41.8
3.7
0.0006
0.405
0 to 500 mV
0.0361
23.9
28.1
2.5
0.0006
0.207
0 to 200 mV
0.0411
14.4
19.6
1.8
0.0006
0.098
0 to 100 mV
0.0461
11.2
18.1
1.7
0.0006
0.059
1
Averaged measurements assume averaging of 100 single-channel readings.
Each PCI-DAS6031 and PCI-DAS6033 is tested at the factory to assure the board’s overall error does not
exceed accuracy limits described in
.
6-2
Содержание PCI-DAS6031