PIN
Designation
Function
Comment
and maximum-value reset
Tare > 100ms in ClickRClackR menu: "Up", goes to
next menu entry.
17
Scale
Scale > 2s: Scaling of the output signal to the
currently effective signal at the input. Default: set at
100% of the output signal. The autoscale level can
be configured to values other than 100% in the
ClickRClackR menu and via the service interface.
Scale > 5s when power switched on: Activate the
ClickRClackR menu
Scale > 100ms in ClickRClackR menu: "Enter",
executes the current menu entry.
16
TEDS
(Peak value)
The connection for 1-wire-EEPROMs functions with
3.3V instead of 5V and includes a 1.5 kR
pull-up resistor of 3.3V. The EEPROMs (e.g.
, DS2430A, DS28EC20) of Maxim/Dallas
are 3.3V compatible.
The software supports TEDS sensors with the
Bridge Sensor ID 33 and
Strain-Gage ID 35 templates.
LED
Status indicator, with signals including "TEDS read",
"Parameter active", "Parameter set", "Gradient
indicator".
max. 4mA, 200 ohm
series resistor;
Temperature
sensor
Type TMP102, -40°C ...+125°C, ±3°C;
GSV-6L
Description
The measurement amplifier GSV-6L adds a configurable analogue output to the GSV-
6CPU. The GSV-6L is designed to be built into sensors and fits into the most confined
housing enclosures, such as connector housings of the "GSV-6K” type.
Even after it is enclosed, there is complete scope to configure all the properties of the
GSV-6L via two "Tare" and "Scale" control cables.
The calibration data from the electronic data sheet of the sensor is automatically accepted
via a TEDS input. The analogue output is then automatically adjusted to the increase set in
TEDS. Dimensions and terminal assignment
ME-Meßsysteme GmbH, Neuendorfstr. 18a, DE-16761 Hennigsdorf
6
Tel +49 (0)3302 78620 60, Fax +49 (0)3302 78620 69, [email protected], www.me-systeme.de
Figure 4: GSV-6L