Technical data GSV-6L / GSV-6K
Accuracy
Accuracy class
0,1
%
Resolution
16
Bit
Strain gauge inputs for full bridge
60 ... 20000
Ohm
Input sensitivity (FS)
0.1 ... 8 (configurable)
mV/V
Common-mode rejection ratio
DC-60Hz
5 kHz
110
100
dB
dB
Noise amplitude "Pk-Pk"
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
20 µV/V (2 mV/V / 10000)
31 µV/V (2 mV/V / 6500)
80 µV/V (2 mV/V / 2500)
140 µV/V ( 2 mV/V / 1400)
240 µV/V ( 2 mV/V / 840)
Zero point drift (at 2 mV/V FS)
<0.05 %FS/10K (1 µV/V / 10K)
Sensitivity drift (at 2 mV/V FS)
<0.01%RD/10K
Measuring frequency
Updating of analogue output
10 ...25000
Hz
Sampling frequency
50
kHz
Adjustment range of averaging filter
0.1 ...25000
Hz
Outputs
Bridge supply voltage
Current carrying capacity
Short-circuit resistance
3.0
60
Yes
Volt
mA
Analogue output
Resolution
Output resistance
Max. output current
Configurable
13 bit
0.12
10
Ohm
mA
Threshold value output,
Current carrying capacity:
Open drain
200 mA, 30V
Interfaces
TEDS
Voltage supply
Nominal voltage
Current consumption
Min. supply voltage
Max. supply voltage
12V … 24V
22
9
28
V
mA
V
V
Operating temperature
ME-Meßsysteme GmbH, Neuendorfstr. 18a, DE-16761 Hennigsdorf
Tel +49 (0)3302 78620 60, Fax +49 (0)3302 78620 69, [email protected], www.me-systeme.de