Pin Configuration
15 pole terminal block for the sensor connection
Figure 1: Terminal block 15pole and 5 pole in GSV-2LS, GSV-2AS, GSV-2ASD and GSV-2FSD
Pin
Name
Description
Tip
1
GND
B
Supply ground
Connection for cable shield
2
+Us
Positive bridge sensor supply
3
+U
F
Positive sens for 6-wire technology
4
+U
D
Positive differential input
5
-U
D
Negative differential input
6
-U
F
Negative sens for 6-wire technology
7
-Us
Negative bridge sensor supply
The signals at terminals 2..7 are
acquired by software. Connection of
sensors in 4- or 6-wire technology.
Bridging bUs and +U
F
as
well as between -U
S
and -U
F
are not
necessary.
8
U
E
Analogue Input 0..10V
9
U
A
Analogue Output
±5 V / option. 4...20mA
10
GND
A
Ground analogue In/Output
Analogue input between U
E
and
GND
A.
Analogue output between U
A
and GND
A
11
SW1 ²
Threshold output No. 1
Open Collector (Open Emitter alternatively),
galvanically isolated
12
Tara
1
Tare input
Affects digital serial and analogue output
13
SW2
2
Threshold output No. 2
See Tip for SW1
14
U
B
Supply voltage 12..24 VDC
15
GND
B
Supply ground
Connection for supply voltage
Table 1: 15pole terminal block in GSV-2LS, GSV-2AS, GSV-2ASD and GSV-2FSD
1
In case of voltages above 3.4 V at this terminal, a tare adjustment is initiated. Offset adjustment is then done in the analog part of
the GSV. In addition the digital output is also set to zero.
2
Threshold values are programmed via the RS 232 port.
ME-Meßsysteme GmbH, Neuendorfstr. 18a, D-16761 Hennigsdorf
4
Phone 03302 78620 60, Fax 03302 78620 69,
,
www.me-systeme.de/en