sign, 6 numbers, decimal point, free space, unit, CR, LF
for example
+1.2345 kgCRLF
Attention: If the unit is switched off (with the command number 15, Set unit, code 7), the
output frame ends with „space“ and CRLF:
+1.2345 CRLF
Display settings
With the binary coded data protocol the measured values are transmitted standardized to
+ 1.
The values sown in the LC display and in the program are a result of the scaling factor
multiplied by the measured value. This scaling factor can be set by the command “Set
Norm“ or with the configuration program.
This is the formula to calculate the scaling factor :
scaling factor = input sensitivity / rated output * normal load.
Example:
normal load of the load cell: 100kg
rated output of the load cell: 2 mV/V
Input sensitivity of the strain gage amplifier: 2 mV/V
==> scaling factor = 100
Output of the register data
After requesting the data, the answer frame sent begins with a semicolon as a prefix.
From 2 up to 8 data bytes are getting transmitted in the frame depending on the size of
the register. You receive the following format:
for 3 Bytes:
;
(ASCII: 59d)
HByte
MByte
LByte
for 2 Bytes:
;
(ASCII: 59d)
HByte
LByte
Commands to the GSV-2
Commands to the GSV have the following format:
Command
number
P1
P2
P3
P4
The Command number may be followed by up to four parameters P1…P4, depending on
the command (see the commands description below).