NIPE Monitor V1 - Continuous monitoring system of comfort
.
/
.
discomfort for newborn infants
MD/PRD/IN16.NIPEV1 v.10 - 12 MAR 2020
34
Electrical Safety: Class 1 conforms to IEC 60601-1
Weight of the monitor with collet fixture (without accessories) = 2.02kg
Dimensions of the monitor with the collet fixture (cm): 27 (width) x 24.7 (height) x 14.5
(depth)
Screen size: 8.4’’, resolution 800
x
600
The efficacy and security of the products are warranted during the lifetime of the products (5
years from the manufacturing date).
Material :
-
Panel Pc: manufacturer IEI (product reference : AFL-08A-N270)
-
Power supply adaptator AC/DC SL POWER - ME40A1203F01
-
Acquisition card: Mdoloris Medical Systems reference: BA-NIPE-V1
Degree of protection against electric shock of the system:
Class 1: equipment in which protection against electric shock does not rely on basic insulation
only, but which includes an additional safety precaution. Means are provided for the
connection of the equipment to the protective earth conductor in the fixed wiring of the
installation in such a way that accessible metal parts cannot become live in the event of a
failure of the basic insulation.
Protection against the ingress of liquids:
Be sure to install the monitor outside any area with a risk of blood or liquids ingress.
System functioning:
Continuous: operation under normal load for a normal period without exceeding the specified
limits of temperature.
Electrosurgery Interference/Defibrillation/Electromagnetic disturbance:
The NIPE Monitor V1 is comply with standard IEC 60601-1-2: 2014 (Ed.4)
The NIPE Monitor V1 is compatible with the use of the HR surgery device.
The equipment returns to the previous operating mode within 10 seconds without loss of any
stored data. Measurement accuracy may be temporarily decreased while performing electro-
surgery or defibrillation. This does not affect patient or equipment safety.