Appendix C: LM-1 error codes
Error 1
Heater circuit shorted
Error 2
Heater circuit open
Error 3
Pump cell circuit shorted
Error 4
Pump cell circuit open
Error 5
Reference cell circuit shorted
Error 6
Reference cell circuit open
Error 7
General System error (typically a software error).
Error 8
Sensor Timing error (typically a damaged sensor).
Error 9
Supply Voltage too low