CUBEScan
TM
BioCon-900S Operator’s Manual
Page 25
Figure 24 Positioning Bladder
Figure 25 Probe Direction
1)
Tap the
Scan
icon(
) on the Information screen or press the probe
Scan
Button
. The scanning
process begins.
2)
The progress bar on the screen appears. It takes 3 seconds to complete the normal scan. Hold the
probe steady while scanning. Movement of the probe during scanning reduces the accuracy of
measurement.
3)
When you hear the beeps, the scan is complete.
Figure 26 Normal Scan Process
5.5.2
PRE-SCAN
The Pre-scan is executed only when Pre-scan setup is ‘ON(
)’ on the Setup screen. When Pre-scan is
‘Off( )’, a normal scan is executed. See the section “7.7 PRE-SCAN SETTING”
1)
The patient should be lying in a supine position and lift up the patient garment exposing the abdominal
region from the pubis to the navel.
2)
Place an ample amount of gel without air bubbles to the probe head. In order to assure optimal
transmission of energy between the patient and the probe, a conductive gel must be applied on the
probe head.
3)
Place the probe head on the patient’s abdomen with the screen facing upwards and aim it to the location
where the bladder is expected to be above 3~4cm the pubic bone.
4)
Ensure that the probe head is pointing toward the patient’s foot.