6

Infrared Tympanic Electronic Thermometer

Pulse Timer Mode

 

1. 

Press and hold the timer button to enter Timer mode.  Press again to start the timer.  The timer will 

 

 

run from 0 to 60 seconds.  

 

2. 

The thermometer will issue a single beep at 15 seconds, a dual beep at 30 seconds, a triple beep at 45 

 

 

seconds, and four beeps at 60 seconds. 

 

3. 

Press the timer button at any point during the timer function to return the thermometer to “off“ mode.

 

4. 

At the end of the 60 seconds, the thermometer will wait two seconds and then enter sleep mode.

 

5. 

Return the thermometer to base for storage. 

Thermometer Display Icons and Alarms 

Thermometer operation is communicated to the user via the LCD display and an audible device contained inside 

the thermometer case. After the probe cover is installed or the batteries have been changed, the thermometer will 

perform a system reset.  The thermometer will perform an internal test to verify that the system components are 

functioning properly.    

Alarm Condition 

 

 

 

      Display Mode

 

 

 

 

 

˚C 

EAR 

Patient temperature above specified range 

 

 

 

Patient temperature below specified range 

 

 

 

 

Ambient Temperature above specified range 
 

Ambient Temperature below specified range 
 

Low Battery  
 

Dead Battery

˚C 

EAR 

EAR 

EAR 

˚C 

EAR 

The LCD displays the low battery icon.  The low battery icon will remain on 
until the battery is replaced or until the dead battery icon is displayed.  After 
the low battery icon is displayed, the user may take approximately 100 
temperatures until the dead battery icon is displayed.

The LCD displays the dead battery condition.  When any button is pressed, the 
battery icon will flash several times and then the LCD will turn OFF.  After the 
battery icon is displayed, the batteries must be changed before the thermometer 
can be used.

Содержание LUMEON 3069

Страница 1: ...n Scale Dimensions Revisions Rev Description Date K PMS 110Yellow Meghan Little 3069_Masterpage Tympanic Thermometers 1 1 Proportion As Provided Final Final Draft 09 03 09 PMS 110Yellow K INFRARED TYM...

Страница 2: ...5 Temperature Display Toggle 5 Sleep Mode 5 Pulse Timer Mode 6 Thermometer Display Icons and Alarms 6 Biotech Mode 7 Section VI Preventative Maintenance 9 Section VII Cleaning 9 Section VIII Battery...

Страница 3: ...Features Very fast temperature acquisition that meets CEN and ASTM standards see Section XI Peak Select System Temperature measurement range of 33 0 C to 42 0 C 91 4 F to 107 6 F Once temperature has...

Страница 4: ...mperatures during cold weather conditions allow the patient to equilibrate to room temperature before use Excessive eardrum scarring may cause lowered temperature readings Under normal conditions ear...

Страница 5: ...Thermometer probe covers could jeopardize the accuracy of the Infrared Tympanic Electronic Thermometer Check your probe cover carton for the LumeonTM logo to be certain that your thermometer will deli...

Страница 6: ...PSS World Medical Inc will result in erroneous temperature measurement In order to aid in infection prevention and control always install a new probe cover prior to taking a temperature The probe cov...

Страница 7: ...age Temperature Recall After a temperature has been acquired the thermometer will enter off mode after approximately 10 seconds Recall the temperature by pressing and releasing the scan button or by p...

Страница 8: ...talled or the batteries have been changed the thermometer will perform a system reset The thermometer will perform an internal test to verify that the system components are functioning properly Alarm...

Страница 9: ...ory calibration parameters are also stored in nonvolatile memory The factory default settings are shown below Temperature mode C unlocked Site mode Ear Site text On The biotech mode is accessed by pre...

Страница 10: ...8 Infrared Tympanic Electronic Thermometer Temperature mode Option Display C unlocked Locked C Locked F F unlocked C C F F Site mode Option Display Oral Core Axillary Rectal Ear ORL CORE AX REC EAR...

Страница 11: ...and or smudges for proper operation A mild detergent may be added to the water Do not use cleaners such as Spray Nine Phisohex Hibiclens or Vesta Syde disinfectants as they may result in damage to the...

Страница 12: ...meter equilibrate in the room for 20 minutes before using For all other system errors reset the thermometer by installing a probe cover If the system error does not clear send the thermometer to the f...

Страница 13: ...ondensing Storage Temperature Range 25 C to 55 C 13 F to 131 F up to 95 RH non condensing If the unit is stored at extremes it is recommended that the unit be checked on the field calibration checker...

Страница 14: ...ry to return a unit for repair please contact PSS World Medical customer information desk at 800 777 4908 Parts Listing To order repair parts contact our customer service center or your sales represen...

Страница 15: ...abeling Comments or Questions about Lumeon Products 800 777 4908 Manufactured for PSS World Medical Inc 4345 Southpoint Blvd Jacksonville FL 32216 Made in the USA www myselectonline com One or more U...

Отзывы: