Infrared Tympanic Electronic Thermometer

Section IX — Mounting Instructions

If using the wall or cart mount options (sold separately) for the base unit, consult the installation sheet provided with the 

respective system.

Section X  — Troubleshooting

If the Infrared Tympanic Electronic Thermometer is not functioning properly, check the following items:  

Symptom                       

 

 

Action

   

 

 

 

 

Temperature reading unusually high 

Check the probe cover for tears or gaps

Temperature reading unusually low 

Check the probe cover and thermometer tip for 

obstructions. Check the patient ear canal for obstructions.

Low battery indicator lit 

Replace battery

   

Dead battery indicator lit  

Replace battery

Display blank 

Replace battery

   

 

System error displayed  

If system error “4”, then let the thermometer equilibrate in 

the room for 20 minutes before using. For all other system 

errors, reset the thermometer by installing a probe cover.  

If the system error does not clear, send the thermometer to 

the facility Biomedical Engineering department or contact 

the factory for available service options. The service 

information is located in Section XII, Customer Service.

The Infrared Tympanic Electronic Thermometer advisory and alarm conditions are described in the Thermometer 

Display Icons and Alarms subheading of Section V, Instructions for Use.

Section XI — Specifications 

Clinical accuracy characteristics and procedures are available from the manufacturer on request. To verify 

accuracy, use a certified blackbody as specified in EN 12470-5-2003, Annex C.

Calibrated Accuracy Limits:  

Ambient Temperature  

Target Temperature  

Accuracy

25°C  

36.7°C to 38.9°C 

± 0.1°C 

(77°F) 

(98.1°F to 102°F) 

(± 0.2°F)

16°C to 33°C 

33°C to 42°C   

± 0.2°C  

 

(60.8°F to 91.4°F) 

(91.4°F to 107.6°F) 

(± 0.4°F)

Содержание LUMEON 3069

Страница 1: ...n Scale Dimensions Revisions Rev Description Date K PMS 110Yellow Meghan Little 3069_Masterpage Tympanic Thermometers 1 1 Proportion As Provided Final Final Draft 09 03 09 PMS 110Yellow K INFRARED TYM...

Страница 2: ...5 Temperature Display Toggle 5 Sleep Mode 5 Pulse Timer Mode 6 Thermometer Display Icons and Alarms 6 Biotech Mode 7 Section VI Preventative Maintenance 9 Section VII Cleaning 9 Section VIII Battery...

Страница 3: ...Features Very fast temperature acquisition that meets CEN and ASTM standards see Section XI Peak Select System Temperature measurement range of 33 0 C to 42 0 C 91 4 F to 107 6 F Once temperature has...

Страница 4: ...mperatures during cold weather conditions allow the patient to equilibrate to room temperature before use Excessive eardrum scarring may cause lowered temperature readings Under normal conditions ear...

Страница 5: ...Thermometer probe covers could jeopardize the accuracy of the Infrared Tympanic Electronic Thermometer Check your probe cover carton for the LumeonTM logo to be certain that your thermometer will deli...

Страница 6: ...PSS World Medical Inc will result in erroneous temperature measurement In order to aid in infection prevention and control always install a new probe cover prior to taking a temperature The probe cov...

Страница 7: ...age Temperature Recall After a temperature has been acquired the thermometer will enter off mode after approximately 10 seconds Recall the temperature by pressing and releasing the scan button or by p...

Страница 8: ...talled or the batteries have been changed the thermometer will perform a system reset The thermometer will perform an internal test to verify that the system components are functioning properly Alarm...

Страница 9: ...ory calibration parameters are also stored in nonvolatile memory The factory default settings are shown below Temperature mode C unlocked Site mode Ear Site text On The biotech mode is accessed by pre...

Страница 10: ...8 Infrared Tympanic Electronic Thermometer Temperature mode Option Display C unlocked Locked C Locked F F unlocked C C F F Site mode Option Display Oral Core Axillary Rectal Ear ORL CORE AX REC EAR...

Страница 11: ...and or smudges for proper operation A mild detergent may be added to the water Do not use cleaners such as Spray Nine Phisohex Hibiclens or Vesta Syde disinfectants as they may result in damage to the...

Страница 12: ...meter equilibrate in the room for 20 minutes before using For all other system errors reset the thermometer by installing a probe cover If the system error does not clear send the thermometer to the f...

Страница 13: ...ondensing Storage Temperature Range 25 C to 55 C 13 F to 131 F up to 95 RH non condensing If the unit is stored at extremes it is recommended that the unit be checked on the field calibration checker...

Страница 14: ...ry to return a unit for repair please contact PSS World Medical customer information desk at 800 777 4908 Parts Listing To order repair parts contact our customer service center or your sales represen...

Страница 15: ...abeling Comments or Questions about Lumeon Products 800 777 4908 Manufactured for PSS World Medical Inc 4345 Southpoint Blvd Jacksonville FL 32216 Made in the USA www myselectonline com One or more U...

Отзывы: