meaning the DC electronic load is waiting for test initiation or remainingin the suspended mode
respectively. Please retrigger it so ttesting continues onward.
After each test is completed, the lower right corner of the VFD display will show either
PASS
or
FAIL
.
If failure occurs, the buzzer will sound. At this point, initiate the next trigger or press any key to free the load
from the pass or fail indication
Once the automated test is finished, press the
▲
and
▼
keys to initiate a manually operated test
mode. With each key press of the
▲
or the
▼
keys one time, the DC electronic load will resume the test of
the previous step or the next step in the sequence. Users can observe the actual state of every step. When
the ON/OFF key is pressed or a trigger is initiated, the DC electronic load will automatically exit from the
manuallyoperated test mode and return to the automatic test mode again.
4.3 Input Control
4.3.1 Short Circuit Operation (SHORT)
The DC Electronic
Load can simulate a short circuit at the input end by turning the load on with
full-scale current. The short circuit function can be toggled on/off at the front panel by pressing the
Shift+9(Short) keys
.
Short circuit operation does not influence the current value set. When the short
circuit operation is in
OFF
state, the Load will return to its original setting state.
The actual current value that the load consumes in short circuit operation depends on the active
working mode and current range of the load.. In CC, CW and CR modes, the maximum short-circuit current
value is 1.2 times the current range. In CV mode, short-circuit operation is identical to setting constant
voltage to 0V.
4.3.2 Input On/Off Operation
When the load input state is
ON
, pressing the
On/Off
key will change the input state to
OFF
. Now top
right corner of the VFD display shows
OFF
. When the load input state is
OFF
, pressing the
On/Off
key will
change the input state to
ON
state. Now the upper right corner of the two line display shows
SHORT
indicating the current working state.
4.4 Electronic Load Operation Range
The DC
Electronic load operatesin the following range of Rated Current, Rated voltage and Rated
Power. Please refer to the Diagram 4-13 and Diagram 4-14.