E02:
Oxygen measurement out of range Low (
≤
-2.0% calculated by algorithm).
Recommended Action: Verify that the UltraMaxO
2
is being used in the correct
mode (Concentrator or Calibration Verification mode). If error code repeats;
perform a calibration verification per section 2.3 of this manual. If error code
repeats again; contact Customer Service.
•
E03:
Device memory corrupt or missing.
Recommended Action: Return the UltraMaxO
2
to the manufacturer for factory
repair.
•
E04:
Signal reading not stable.
Recommended Action: Return the UltraMaxO
2
to the manufacturer for factory
repair.
•
E05:
Pressure measurement out of Range Hi (
≥
50 PSI).
Recommended Action: Check the pressure on a known gas source pressure. If
error code repeats; contact Customer Service.
•
E06:
Outside of operating temperature Hi (
≥
40° C).
Recommended action: The UltraMaxO
2
is too hot, cool the device closer to room
temperature before use.
•
E07:
Outside of operating temperature Low (
≤
15° C).
Recommended action: The UltraMaxO
2
is too cold, warm the device closer to room
temperature before use.
•
E08:
Device self check found error.
Recommended Action: Remove and replace the batteries. If error code repeats;
return the UltraMaxO
2
to the manufacturer for factory repair.
5.0 CHANGING THE BATTERIES
Batteries should be changed when the battery icon illuminates. The battery icon will
remain lit until the batteries are changed. If the battery power level is too low the
UltraMaxO
2
will not power on until the batteries are changed.
5.1 Battery Replacement Procedure
1. Using a #1 Phillips screwdriver loosen the battery door screw and remove the
battery door.
2. Remove the batteries.
3. Insert new batteries ensuring correct placement.
DO NOT
use rechargeable
batteries.
4. Replace the battery door and tighten the battery door screw.
5. If the UltraMaxO
2
does not power on when done verify the batteries are installed
correctly and that the batteries are fresh.
6.0 CLEANING AND MAINTENANCE
•
Use caution to prevent any fluid from entering the UltraMaxO
2
.
•
DO NOT
soak or immerse the UltraMaxO
2
in fluid.
•
DO NOT
autoclave or expose the UltraMaxO
2
to ethylene oxide sterilization.
6.1 Cleaning
•
Wipe down the exterior surfaces of the UltraMaxO
2
with a moist cloth and mild
hand or dish soap (pH 6-8).